应用于胆管的冷等离子体内窥镜:针对胆管癌治疗在仿生及猪模型上的可行性风险评估
Cold plasma endoscopy applied to biliary ducts: feasibility risk assessment on human-like and porcine models for the treatment of cholangiocarcinoma
AI总结:
本研究评估了冷等离子体内窥镜治疗胆管癌的可行性与安全性,通过8-10 kV脉冲驱动转移型导管,在人工胆管和猪模型中验证无电热风险且功率约100 mW,并证实其在体外人CCA模型中具抗肿瘤效应。
摘要:
过去十二年来,冷等离子体因其抗肿瘤特性,已被定位为治疗癌症的突破性技术。针对预后极差的癌症,定制化等离子体源和个性化治疗方案的创新显得尤为关键。胆管癌(CCA)即属此类,其冷等离子体治疗需要创新导管和内窥镜设备以实现局部治疗。在开展临床试验前,必须从患者与医生的安全性以及治疗功效两方面评估冷等离子体内窥镜的性能与局限性。本研究旨在实现这些目标:使用一种转移型等离子体导管,由8-10 kV、5-10 kHz的重复脉冲驱动,使引导流光能够传输至少2米的长度。第一步,将导管直接插入一个重现胆管系统拓扑结构和电响应的人工模型中。该模型验证了该技术的技术可行性,并证明不存在电学和热学风险。第二步,将导管插入十二指肠镜,整体应用于猪解剖模型。在穿过食道、胃和十二指肠后,十二指肠镜的远端被置于十二指肠乳头入口处,使导管能顺利进入胆总管及肝总管。值得注意的是,沉积的电功率值约为100 mW,尤其因为电流值至少高出10倍。IEC医疗器械标准进一步证明并巩固了不存在电学和热学风险的结论。第三步,我们证明冷等离子体导管能够在体外人类CCA实验模型上诱导抗肿瘤效应。
英文摘要:
For the past twelve years, cold plasmas have been positioned as a breakthrough technology for treating cancer thanks to their antitumor properties. The innovation of ad hoc plasma sources and personalized protocols appears crucial to treat cancers with a very poor prognosis. This is the case for cholangiocarcinoma (CCA) , whose treatment with cold plasma requires the innovation of catheters and endoscopic devices for local therapies. Before conducting clinical trials, the performances and limitations of cold plasma endoscopy must be evaluated in terms of safety for both the patient and clinician as well as in terms of therapeutic efficacy. These objectives are pursued in the present work, in which a transferred plasma catheter is used, powered by 8-10 kV, 5-10 kHz repeated pulses, allowing guided streamers to be transferred over lengths of at least 2 m. In a first step, the catheter is directly inserted into an artificial model reproducing the topography and the electrical response of the biliary tree. This model allows to validate the technical feasibility of the technology and to demonstrate the absence of electrical and thermal risks. In a second step, the catheter is inserted into the duodenoscope, the whole being applied to a porcine anatomical model. After passing through the esophagus, stomach and duodenum, the distal part of the duodenoscope is placed at the entrance to the papilla so that the catheter can easily enter the choledoc and then the common bile duct. Interestingly, the electrical power values deposited are of the order of 100 mW especially because the current values are at least 10 times higher. The absence of electrical and thermal risks is demonstrated and consolidated by the IEC standards for medical devices. In a third time, we demonstrate that the cold plasma catheter can induce antitumor effects on in vitro experimental models of human CCA.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。