Non-invasive urinary bladder volume estimation with artefact-suppressed bio-impedance measurements

Kanika Dheman, Stefan Walser, Philipp Mayer, Manuel Eggimann, Marko Kozomara, Denise Franke, Thomas Hermanns, Hugo Sax, Simone Schürle, Michele Magno

AI总结:

本文提出ANUVIS可穿戴生物阻抗系统与深度学习伪影抑制算法,实现连续、无创、免手持的膀胱容量定量估计,精度媲美便携式超声。

摘要:

尿量是衡量肾脏健康的重要参数。当前的监测方法包括人工书面记录、侵入性导尿,或由高技能人员进行超声测量。导尿具有较高的感染风险,而间歇性超声测量和人工记录耗时,并可能错过肾功能异常的早期迹象。生物阻抗(BI)测量可作为在体内测量尿量的一种无创替代方案。然而,有限的鲁棒性阻碍了其临床转化。本文提出一种基于深度学习的算法,该算法处理下腹部的局部BI并抑制伪影,从而定量、无创且无需持续额外人员在场即可测量膀胱容量。研究使用名为ANUVIS的四电极BI可穿戴系统,从三名健康受试者采集连续膀胱容量数据以证明操作可行性,同时以尿动力学测试(n=6)和尿流率测定测试(n=8)的临床金标准提供真值。研究推导了电极放置的优化位置,以及BI随膀胱容量变化的模型。全膀胱容量估计和残余容量估计的平均误差为-29 ± 87.6 ml,因此可与商用便携式超声设备相媲美(Bland Altman分析显示偏倚为-5.2 ml,LoA在119.7 ml至-130.1 ml之间),同时还具有免手持、无创和连续膀胱容量估计的额外优势。可穿戴BI传感器节点与所提出算法的结合,为当前标准护理提供了一种有吸引力的替代方案,并可能在深入了解肾功能方面带来益处。

英文摘要:

Urine output is a vital parameter to gauge kidney health. Current monitoring methods include manually written records, invasive urinary catheterization or ultrasound measurements performed by highly skilled personnel. Catheterization bears high risks of infection while intermittent ultrasound measures and manual recording are time consuming and might miss early signs of kidney malfunction. Bioimpedance (BI) measurements may serve as a non-invasive alternative for measuring urine volume in vivo. However, limited robustness have prevented its clinical translation. Here, a deep learning-based algorithm is presented that processes the local BI of the lower abdomen and suppresses artefacts to measure the bladder volume quantitatively, non-invasively and without the continuous need for additional personnel. A tetrapolar BI wearable system called ANUVIS was used to collect continuous bladder volume data from three healthy subjects to demonstrate feasibility of operation, while clinical gold standards of urodynamic (n=6) and uroflowmetry tests (n=8) provided the ground truth. Optimized location for electrode placement and a model for the change in BI with changing bladder volume is deduced. The average error for full bladder volume estimation and for residual volume estimation was -29 +/-87.6 ml, thus, comparable to commercial portable ultrasound devices (Bland Altman analysis showed a bias of -5.2 ml with LoA between 119.7 ml to -130.1 ml), while providing the additional benefit of hands-free, non-invasive, and continuous bladder volume estimation. The combination of the wearable BI sensor node and the presented algorithm provides an attractive alternative to current standard of care with potential benefits in providing insights into kidney function.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑