头部传感器运动学的频率内容与滤波:一种实现基于现场的研究间比较的方法
Frequency content and filtering of head sensor kinematics: A method to enable field-based inter-study comparisons
AI总结:
本研究开发了一种伪影衰减滤波方法,通过分析实验室和现场头部运动学信号的频率特征,提高信噪比并实现不同可穿戴传感器系统间的现场研究比较。
摘要:
可穿戴头部传感器系统使用不同的运动学信号处理方法,这限制了基于现场的研究间比较,尤其是当信号中存在伪影时。本研究的目的是评估来自头部撞击重建实验室和现场环境的头部运动学信号的频率内容和特征,以开发一种伪影衰减滤波方法(伪影衰减方法)。在测试假人头部模型上进行了实验室撞击(n=72),峰值加速度范围为25-150 g,并为126名精英级橄榄球联合会球员配备了带仪器的护齿器(iMG),最多进行四场比赛。利用实验室撞击和现场HAE(n=5694)的功率谱密度(PSD)特征,如第95百分位累积和PSD频率,来开发伪影衰减方法。通过信噪比(SNR)和混合线性效应模型,将伪影衰减方法与另外两种常见的滤波方法(四阶(2x2极点)、零滞后Butterworth滤波器,截止频率为200 Hz(-6 dB)(Butterworth-200Hz)和CFC180滤波器)分别针对实验室和现场事件进行比较。伪影衰减方法产生了比Butterworth-200Hz和CFC180滤波器更高的总体SNR,且在实验室测试的量级范围内,现场峰值线性加速度(PLA)和峰值角加速度(PAA)值也更高。CFC180滤波器的中位PLA和PAA高于Butterworth-200Hz(p<0.01)和伪影衰减方法(p<0.01),报告的值高达294 g和31.2 krad/s2。伪影衰减方法可应用于所有具有足够采样率的商用iMG运动学信号,以实现基于现场的研究间比较。
英文摘要:
Wearable head sensor systems use different kinematic signal processing approaches which limits field-based inter-study comparisons, especially when artefacts are present in the signal. The aim of this study is to assess the frequency content and characteristics of head kinematic signals from head impact reconstruction laboratory and field-based environments to develop an artefact attenuation filtering method (artefact attenuation method). Laboratory impacts (n=72) on a test-dummy headform ranging from 25-150 g were conducted and 126 elite-level rugby union players were equipped with instrumented mouthguards (iMG) for up to four matches. Power spectral density (PSD) characteristics of the laboratory impacts and on-field HAE (n=5694) such as the 95th percentile cumulative sum PSD frequency were utilised to develop the artefact attenuation method. The artefact attenuation method was compared to two other common filtering approaches (Fourth order (2x2 pole), zero-lag Butterworth filter with 200 Hz (-6 dB) cut-off frequency (Butterworth-200Hz) and CFC180 filter) through signal-to-noise ratio (SNR) and mixed linear effects models for laboratory and on-field events, respectively. The artefact attenuation method produced an overall higher SNR than the Butterworth-200Hz and CFC180 filter and on-field peak linear acceleration (PLA) and peak angular acceleration (PAA) values within the magnitude range tested in the laboratory. Median PLA and PAA were higher for the CFC180 filter than the Butterworth-200Hz (p<0.01) and artefact attenuation method (p<0.01), reporting values as high as 294 g and 31.2 krad/s2. The artefact attenuation method can be applied to all commercially available iMG kinematic signals with adequate sample rates to enable field-based inter-study comparisons.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。