一种新型功耗优化的CMOS sEMG设备,集成超低噪声ConvNet(VGG16),用于生物医学应用
A Novel Power-optimized CMOS sEMG Device with Ultra Low-noise integrated with ConvNet (VGG16) for Biomedical Applications
AI总结:
本研究开发了一种功耗优化的CMOS sEMG设备,集成VGG16网络,显著降低噪声并提高数据处理效率,在肌张力障碍诊断和思维控制假肢应用中达到95.8%的准确率。
摘要:
用于测量肌肉和大脑活动的针式生物电位传感器需要侵入性手术靶向肌肉神经再支配(TMR)以及繁琐的维护过程,而表面生物电位传感器缺乏清晰的生物信号读取(信号干扰)。在本研究中,开发了一种新型功耗优化的互补金属氧化物半导体(CMOS)表面肌电图(sEMG)设备,以提高所捕获生物信号的效率和质量,用于生物医学应用:神经系统疾病(肌张力障碍)的早期诊断,以及一种与人类日常活动兼容的新型思维控制假肢。一种由基于CMOS运算放大器的PIC16F877A 8位CMOS闪存微控制器组成的新型sEMG被用于最小化功耗和数据处理时间。sEMG电路通过开发的模拟滤波器以及经由快速傅里叶变换(FFT)、Z变换和差分方程实现的无限脉冲响应(IIR)数字滤波器来实现。分析显示,根据数值均方根误差(RMSE),使用开发的数字滤波器相比模拟滤波器,记录的EMG信号噪声降低了169.2%。此外,数字IIR在两个阶段进行了测试:算法阶段和实际世界阶段。结果显示,IIR的算法(MATLAB)和实际世界RMSE分别为0.03616和0.05224。在EMG信号分析中,数据处理时间显著提高了20.8%。优化了VGG、AlexNet和ResNet ConvNet,并在15个公共EEG(62电极)和18个受试者的观察EMG数据上进行了训练和测试。结果表明,VGG16-1D的准确率高出98.43%。在实际测试中,16名受试者(6名截肢者-10名肌张力障碍患者)的准确率为95.8±4.6%。本研究展示了sEMG的潜力,为生物医学应用铺平了道路。
英文摘要:
The needle bio-potential sensors for measuring muscle and brain activity need invasive surgical targeted muscle reinnervation (TMR) and a demanding process to maintain, but surface bio-potential sensors lack clear bio-signal reading (Signal-Interference). In this research, a novel power-optimized complementary metal-oxide-semiconductor (CMOS) Surface Electromyography (sEMG) is developed to improve the efficiency and quality of captured bio-signal for biomedical application: The early diagnosis of neurological disorders (Dystonia) and a novel compatible mind-controlled prosthetic leg with human daily activities. A novel sEMG composed of CMOS Op-Amp based PIC16F877A 8-bit CMOS Flash-based Microcontroller is utilized to minimize power consumption and data processing time. sEMG Circuit is implemented with developed analog filter along with infinite impulse response (IIR) digital filter via Fast Fourier Transform (FFT), Z-transform, and difference equations. The analysis shows a significant improvement of 169.2% noise-reduction in recorded EMG signal using developed digital filter compared to analog one according to numerical root mean square error (RMSE). Moreover, digital IIR was tested in two stages: algorithmic and real-world. As a result, IIR's algorithmic (MATLAB) and real-world RMSEs were 0.03616 and 0.05224, respectively. A notable advancement of 20.8% in data processing duration in EMG signal analysis. Optimizing VGG, AlexNet, and ResNet ConvNet as trained and tested on 15 public EEG (62-electrode) and 18 subjects' observed EMG data. The results indicate that VGG16-1D is 98.43% higher. During real testing, the accuracy was 95.8 +/- 4.6% for 16 subjects (6 Amputees-10 Dystonia). This study demonstrates the potential for sEMG, paving the way for biomedical applications.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。