超声诊断性能评价用凝胶蜡模:图像均匀性、几何精度和高回声目标直径的评估

A gel wax phantom for performance evaluation in diagnostic ultrasound: assessment of image uniformity, geometric accuracy, and diameter of a hyperechoic target

Debjani Phani, Rajasekhar Konduru Varadarajulu, Arijit Paramanick, Souradip Paul, Raghukumar Paramu, George Zacharia, Shaiju VS, Venugopal Muraleedharan, M. Suh… 展开作者

Debjani Phani, Rajasekhar Konduru Varadarajulu, Arijit Paramanick, Souradip Paul, Raghukumar Paramu, George Zacharia, Shaiju VS, Venugopal Muraleedharan, M. Suheshkumar Singh, Raghuram Kesavan Nair

摘要:

目的:通过将目标嵌入凝胶蜡中来确定图像均匀性、横向和轴向分辨率以及不锈钢圆盘的直径,开发并确认诊断超声(US)扫描仪的体模。材料与方法:测定了凝胶蜡的声学特性,包括超声传播速度(cus)、声阻抗(Z)和衰减系数。使用脉冲回波技术估计cus和衰减系数。Z由样品密度(\r{ho})和cus的乘积获得。使用聚四氟乙烯(PTFE)板构造两个矩形框架,所述矩形框架具有间隔5、10和20 mm距离的孔。将尼龙长丝和SS盘(直径= 16.8mm)穿过框架并适当地放置在熔化的凝胶蜡中以获得正交目标。使用计算机断层扫描(CT)和2-9 MHz US探头测量目标。结果:凝胶蜡的AP为cus=1418 m/s,\r{ho} = 0.87 g/cm 3,Z=1.23 MRayls,衰减系数为0.88 dB/cm/MHz。将目标的US成像结果与其物理尺寸进行比较,并作为扫描仪和探头的基线。体模US图像中距离测量的最大误差为7.1%,62周内体模体积减少了1.8%。结论:凝胶蜡可用于开发可广泛应用的价格合理、高度稳定和可定制的诊断性US体模。

英文摘要:

Purpose: To develop and validate a phantom for diagnostic ultrasound (US) scanners by embedding targets in gel wax to determine the image uniformity, lateral and axial resolution, and diameter of a stainless-steel disc. Materials and Methods: Acoustic property (AP), which includes the velocity of US (cus), acoustic impedance (Z), and attenuation coefficient in gel wax were determined. The cus, and attenuation coefficient were estimated using the pulse-echo technique. Z was obtained from the product of sample density (\r{ho}) and cus. Two rectangular frames using polytetrafluoroethylene (PTFE) sheets with holes separated by 5, 10, and 20 mm distances were constructed. Nylon filaments and SS-disc (diameter = 16.8 mm) were threaded through the frames and suitably placed in melted gel wax to obtain orthogonal targets. The targets were measured using computerized tomography (CT) and a 2-9 MHz US probe. Results: The AP of gel wax were cus=1418 m/s, \r{ho}= 0.87 g/cm3, Z=1.23 MRayls, attenuation coefficient= 0.88 dB/cm/MHz. The results of US imaging of the targets were compared with their physical sizes and served as baselines for the scanner and probe. The maximum error in distance measurement in the phantom US images was 7.1%, and the phantom volume decreased by 1.8% over 62 weeks. Conclusion: Gel wax can be useful in developing affordable, highly stable, and customizable diagnostic US phantoms that can be implemented widely.