StarDICE I:传感器校准台与索尼 IMX411 传感器的绝对光度校准
StarDICE I: sensor calibration bench and absolute photometric calibration of a Sony IMX411 sensor
AI总结:
StarDICE 实验建立从 NIST 光电二极管到 CMOS 传感器的 5 级计量链,实现绝对光度校准,量子效率不确定度低于 0.001 e-/γ,整体归一化不确定度 1%。
摘要:
Ia 型超新星(SNe-Ia)的哈勃图提供了对暗能量本质的宇宙学约束,其精度受限于当前可用分光光度标准的通量校准。StarDICE 实验旨在建立从 NIST 光电二极管到恒星的 5 级计量链,目标精度为 $griz$ 波段的 \\(\SI{1}{mmag}\\)。我们展示了前两个阶段,实现了从 NIST 光电二极管到演示用 \\(\SI{150}{Mpixel}\\) CMOS 传感器(QHYCCD 的 QHY411M 相机中采用的索尼 IMX411ALR)的校准传递。作为副产品,我们提供了该相机的完整表征。我们构建了一个全自动分光光度台以执行校准传递。利用数千张平场图像研究了传感器读出电子学,从中推导出稳定性、高分辨率光子传递曲线以及单个像素增益的估计值。随后,传感器量子效率相对于 NIST 校准的光电二极管进行测量。使用 16 个不同波长的平场扫描来构建传感器响应图。我们展示了在 \\(\SI{387}{nm}\\) 至 \\(\SI{950}{nm}\\) 之间量子效率的统计不确定度低于 \\(\SI{0.001}{e^-/\gamma}\\)。校准台光学系统中的系统不确定度控制在 \\(\SI{1e-3}{e^-/\gamma}\\) 水平。QE 曲线整体归一化的不确定度为 1%。关于相机,我们展示了稳态条件下的稳定性为 \\(\SI{32.5}{ppm}\\)。整个传感器区域的响应均匀性低于 \\(\SI{1}{\percent}\\) RMS。量子效率在可见光大部分范围内保持在 \\(\SI{50}{\percent}\\) 以上,在 \\(\SI{440}{nm}\\) 至 \\(\SI{570}{nm}\\) 之间峰值远高于 \\(\SI{80}{\percent}\\)。检测到 \\(\SI{1}{\percent}\\) 水平的微分非线性。提出了一个简单的 2 参数模型以减轻该效应。
英文摘要:
The Hubble diagram of type-Ia supernovae (SNe-Ia) provides cosmological constraints on the nature of dark energy with an accuracy limited by the flux calibration of currently available spectrophotometric standards. The StarDICE experiment aims at establishing a 5-stage metrology chain from NIST photodiodes to stars, with a targeted accuracy of \SI{1}{mmag} in $griz$ colors. We present the first two stages, resulting in the calibration transfer from NIST photodiodes to a demonstration \SI{150}{Mpixel} CMOS sensor (Sony IMX411ALR as implemented in the QHY411M camera by QHYCCD). As a side-product, we provide full characterization of this camera. A fully automated spectrophotometric bench is built to perform the calibration transfer. The sensor readout electronics is studied using thousands of flat-field images from which we derive stability, high resolution photon transfer curves and estimates of the individual pixel gain. The sensor quantum efficiency is then measured relative to a NIST-calibrated photodiode. Flat-field scans at 16 different wavelengths are used to build maps of the sensor response. We demonstrate statistical uncertainty on quantum efficiency below \SI{0.001}{e^-/γ} between \SI{387}{nm} and \SI{950}{nm}. Systematic uncertainties in the bench optics are controlled at the level of \SI{1e-3}{e^-/γ}. Uncertainty in the overall normalization of the QE curve is 1\%. Regarding the camera we demonstrate stability in steady state conditions at the level of \SI{32.5}{ppm}. Homogeneity in the response is below \SI{1}{\percent} RMS across the entire sensor area. Quantum efficiency stays above \SI{50}{\percent} in most of the visible range, peaking well above \SI{80}{\percent} between \SI{440}{nm} and \SI{570}{nm}. Differential non-linearities at the level of \SI{1}{\percent} are detected. A simple 2-parameter model is proposed to mitigate the effect.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。