经口肌腱驱动关节机器人单元的准静态分析

Quasi-Static Analysis on Transoral Surgical Tendon-Driven Articulated Robot Units

Georgia Institute of Technology(佐治亚理工学院);Asan Medical Center(峨山医院);Asan Institute for Life Sciences(峨山生命科学研究所);University of Ulsan College of Medicine(蔚山大学医学院)

Hojin Seo, Yeoun-Jae Kim, Jaesoon Choi, Youngjin Moon

摘要:

肌腱驱动连续体机器人中的线致动使得能够从一定距离传递力,但是应当理解,当滑轮用于以推拉模式致动两个线缆时,可能出现张力控制问题。本文分析了在准静态平衡中,单个连续体单元的转角、压力以及变量之间的关系。准静态分析的主要目标是输出压力和分析,给定施加的张力。建立了静力平衡条件,并对转角进行了二分法求解。二分法的函数考虑了压力诱导力、摩擦力和重量。当Tl和Ts分别为1N和2N时,θ为17.14°,p为405.6Pa。结果与初步设计指标基本一致,需要进一步的模拟和实验。

英文摘要:

Wire actuation in tendon-driven continuum robots enables the transmission of force from a distance, but it is understood that tension control problems can arise when a pulley is used to actuate two cables in a push-pull mode. This paper analyzes the relationship between angle of rotation, pressure, as well as variables of a single continuum unit in a quasi-static equilibrium. The primary objective of the quasi-static analysis was to output pressure and the analysis, given the tensions applied. Static equilibrium condition was established, and the bisection method was carried out for the angle of rotation. The function for the bisection method considered pressure-induced forces, friction forces, and weight. θ was 17.14°, and p was 405.6 Pa when Tl and Ts were given the values of 1 N and 2 N, respectively. The results seemed to be consistent with the preliminary design specification, calling for further simulations and experiments.