荧光核径迹探测器光漂白复用性研究

Study on the reusability of fluorescent nuclear track detectors using optical bleaching

Abdul Muneem, Junya Yoshida, Hiroyuki Ekawa, Masahiro Hino, Katsuya Hirota, Go Ichikawa, Ayumi Kasagi, Masaaki Kitaguchi, Satoshi Kodaira, Kenji Mishima, Jameel… 展开作者

Abdul Muneem, Junya Yoshida, Hiroyuki Ekawa, Masahiro Hino, Katsuya Hirota, Go Ichikawa, Ayumi Kasagi, Masaaki Kitaguchi, Satoshi Kodaira, Kenji Mishima, Jameel-Un Nabi, Manami Nakagawa, Michio Sakashita, Norihito Saito, Takehiko R. Saito, Satoshi Wada, Nakahiro Yasuda

摘要:

基于Al ${_2}$ O ${_3}$:C,Mg晶体的荧光核径迹探测器(FNTD)是可用于带电粒子和中子的剂量测定和探测的发光探测器。这些探测器可用于成像应用,其中需要相当高的轨道密度,在100 $^2$ $\times$ 100 $μ$ m的区域中大约1个$\times$ $10^4$轨道的数量级。为了研究FNTD成像应用的可重用性,我们提出了一种方法来执行所需的轨道密度条件下的光学漂白。通过7次辐照漂白循环评估可重复使用性。对于辐照,将研究的FNTD暴露于来自$^{241}$ Am放射源的α粒子。通过波长为355 nm的紫外激光进行光学漂白。三个专用区域上的一个单一的FNTD与不同的累积轨道密度和漂白条件进行了研究。在每个辐照-漂白循环之后,计算信噪比以评价FNTD性能。它的结论是,FNTDs可以重复使用至少七次的应用程序中积累的高轨道密度是必需的。

英文摘要:

Fluorescent nuclear track detectors (FNTDs) based on Al${_2}$O${_3}$:C,Mg crystals are luminescent detectors that can be used for dosimetry and detection of charged particles and neutrons. These detectors can be utilised for imaging applications where a reasonably high track density, approximately of the order of 1 $\times$ $10^4$ tracks in an area of 100 $\times$ 100 $μ$m$^2$, is required. To investigate the reusability of FNTDs for imaging applications, we present an approach to perform optical bleaching under the required track density conditions. The reusability was assessed through seven irradiation-bleaching cycles. For the irradiation, the studied FNTD was exposed to alpha-particles from an $^{241}$Am radioactive source. The optical bleaching was performed by means of ultraviolet laser light with a wavelength of 355 nm. Three dedicated regions on a single FNTD with different accumulated track densities and bleaching conditions were investigated. After every irradiation-bleaching cycle, signal-to-noise ratio was calculated to evaluate FNTD performance. It is concluded that FNTDs can be reused at least seven times for applications where accumulation of a high track density is required.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑