通过对RF无屏蔽MRI的同时感测和深度学习预测实现稳健的电磁干扰(EMI)消除
Robust Electromagnetic Interference (EMI) Elimination via Simultaneous Sensing and Deep Learning Prediction for RF Shielding-free MRI
摘要:
现时,核磁共振扫描是在一个全封闭的射频屏蔽室内进行,对安装有严格的要求,亦会对病人造成不必要的不适。我们的目标是开发一种电磁干扰(EMI)消除策略,用于无RF屏蔽或不完全RF屏蔽的MRI。在这项研究中,提出了一种同时感测和深度学习驱动的EMI消除策略,以从采集的MRI信号中建模、预测和去除EMI信号。具体地,在每次MRI扫描期间,利用放置在各种空间位置中的单独的EMI感测线圈来同时对两个窗口内的环境和内部EMI信号进行采样(对于常规MRI信号采集和EMI表征采集两者)。然后,使用EMI表征数据来训练CNN模型,以将由EMI感测线圈检测到的EMI信号与MRI接收线圈中的EMI信号相关联。该模型用于回顾性预测和去除MRI接收线圈在MRI信号采集窗口期间检测到的EMI信号分量。我们实施并证明了这种策略的各种EMI源上的移动超低场0.055 T永磁MRI扫描仪和1.5 T超导磁体MRI扫描仪没有或不完全的RF屏蔽。我们的实验结果表明,该方法是非常有效的和强大的预测和消除各种电磁干扰源从外部环境和内部扫描仪电子设备在0.055 T(2.3 MHz)和1.5 T(64 MHz),产生最终的图像信噪比是使用一个完全封闭的RF屏蔽获得的。我们提出的策略使MRI操作没有或不完整的射频屏蔽,减轻MRI安装和操作要求。它还可能适用于在存在外部和内部EMI或RF源的情况下准确检测或区分RF信号的其他场景。
英文摘要:
At present, MRI scans are performed inside a fully-enclosed RF shielding room, posing stringent installation requirement and unnecessary patient discomfort. We aim to develop an electromagnetic interference (EMI) cancellation strategy for MRI with no or incomplete RF shielding. In this study, a simultaneous sensing and deep learning driven EMI cancellation strategy is presented to model, predict and remove EMI signals from acquired MRI signals. Specifically, during each MRI scan, separate EMI sensing coils placed in various spatial locations are utilized to simultaneously sample environmental and internal EMI signals within two windows (for both conventional MRI signal acquisition and EMI characterization acquisition). Then a CNN model is trained using the EMI characterization data to relate EMI signals detected by EMI sensing coils to EMI signals in MRI receive coil. This model is utilized to retrospectively predict and remove EMI signals components detected by MRI receive coil during the MRI signal acquisition window. We implemented and demonstrated this strategy for various EMI sources on a mobile ultra-low-field 0.055 T permanent magnet MRI scanner and a 1.5 T superconducting magnet MRI scanner with no or incomplete RF shielding. Our experimental results demonstrate that the method is highly effective and robust in predicting and removing various EMI sources from both external environments and internal scanner electronics at both 0.055 T (2.3 MHz) and 1.5 T (64 MHz), producing final image signal-to-noise ratios that are comparable to those obtained using a fully enclosed RF shielding. Our proposed strategy enables MRI operation with no or incomplete RF shielding, alleviating MRI installation and operational requirements. It is also potentially applicable to other scenarios of accurate RF signal detection or discrimination in presence of external and internal EMI or RF sources.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。