锗探测器中电子漂移各向异性温度依赖性及其对电子漂移模型的启示
Temperature Dependence of the Electron-Drift Anisotropy and Implications for the Electron-Drift Model
AI总结:
研究了锗探测器中电子漂移的温度依赖性,提出修改电子漂移模型以提高模拟准确性。
摘要:
锗探测器中的电子漂移建模涉及许多假设。面对数据,这些假设必须重新审视。对n型分段点接触锗探测器的电子漂移的温度依赖性进行了详细研究。该探测器安装在温度可控、电冷却的低温恒温器中。表面事件通过来自$^{133}$Ba源的聚焦81 keV光子诱导。对在不同温度下通过表面扫描收集的脉冲上升时间进行的详细分析被呈现。纵向电子漂移各向异性随着温度升高而降低。一种新的方法被采用,该方法利用由SolidStateDetectors$.$jl模拟确定的指定上升时间窗口,用于分离沿不同轴的电子纵向漂移以量化这一观察。测量的纵向漂移速度的温度依赖性结合标准电子漂移模型(在相关模拟包中广泛使用)导致不合理的预测。一种修改电子漂移模型的初步建议被提出并描述。在SolidStateDetectors$.$jl中首次实现修改模型的结果被展示。它们合理地描述了数据。建议对模型和标准输入值进行一般回顾。
英文摘要:
The electron drift in germanium detectors is modeled making many assumptions. Confronted with data, these assumptions have to be revisited. The temperature dependence of the drift of electrons was studied in detail for an n-type segmented point-contact germanium detector. The detector was mounted in a temperature controlled, electrically cooled cryostat. Surface events were induced with collimated 81 keV photons from a $^{133}$Ba source. A detailed analysis of the rise time of pulses collected in surface scans, performed at different temperatures, is presented. The longitudinal anisotropy of the electron drift decreases with rising temperature. A new approach, making use of designated rise-time windows determined by simulations using SolidStateDetectors$.$jl, was used to isolate the longitudinal drift of electrons along different axes to quantify this observation. The measured temperature dependence of the longitudinal drift velocities combined with the standard electron drift model as widely used in relevant simulation packages results in unphysical predictions. A first suggestion to modify the electron-drift model is motivated and described. The results of a first implementation of the modified model in SolidStateDetectors$.$jl are shown. They describe the data reasonably well. A general review of the model and the standard input values for mobilities is suggested.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。