使用初级调制器和双层探测器的单次激发定量X射线成像

Single-Shot Quantitative X-ray Imaging Using a Primary Modulator and Dual-Layer Detector

Linxi Shi, N. Robert Bennett, Alexander Vezeridis, Nishita Kothary, Adam Wang

摘要:

目的:传统的X射线成像和荧光透视由于包括散射、射束硬化和重叠组织在内的若干挑战而在定量方面具有局限性。在这项工作中,我们提出了单次激发定量成像(SSQI)通过结合使用的初级调制器(PM)和双层(DL)检测器,这使得运动自由双能量(DE)成像与散射校正在一个Single Shot。研究方法:SSQI算法允许使用来自PM编码的四个子测量同时恢复两个材料特定图像和两个散射图像。为了进行验证,我们在桌面系统上设置了SSQI,并对已知厚度的丙烯酸和铜板进行了成像,使用我们的SSQI算法估计了散射,并将两种材料的不同组合的材料分解(MD)与地面实况进行了比较。第二,我们对一个在冠状动脉中含有造影剂的拟人胸部体模进行成像,并比较了有和没有SSQI的MD。最后,为了评估SSQI在动态应用中的效果,我们构建了一个能够对碘造影剂进行动态成像的流动体模。结果:SSQI估计的均方根误差(RMSE)为0.13厘米的丙烯酸和0.04毫米的铜。对于拟人体模,直接MD导致对比度和软组织的错误解释,而SSQI成功区分它们,并将材料特定图像的RMSE降低了38%至92%。对于流动体模,SSQI能够执行准确的动态定量成像,将对比度与背景分离。结论:我们证明了SSQI用于稳健的定量X射线成像的潜力。SSQI的简单性可以使其广泛采用,包括放射摄影和动态成像,如实时图像引导和锥束CT。

英文摘要:

Purpose: Conventional x-ray imaging and fluoroscopy have limitations in quantitation due to several challenges, including scatter, beam hardening, and overlapping tissues. In this work, we propose single-shot quantitative imaging (SSQI) by combining the use of a primary modulator (PM) and dual-layer (DL) detector, which enables motion-free dual-energy (DE) imaging with scatter correction in a single shot. Methods: The SSQI algorithm allows simultaneous recovery of two material-specific images and two scatter images using four sub-measurements from the PM encoding. For validation, we set up SSQI on our tabletop system and imaged acrylic and copper slabs with known thicknesses, estimated scatter with our SSQI algorithm, and compared the material decomposition (MD) for different combinations of the two materials with ground truth. Second, we imaged an anthropomorphic chest phantom containing contrast in the coronary arteries and compared the MD with and without SSQI. Lastly, to evaluate SSQI in dynamic applications, we constructed a flow phantom that enabled dynamic imaging of iodine contrast. Results: The root mean squared error (RMSE) of SSQI estimation was 0.13 cm for acrylic and 0.04 mm for copper. For the anthropomorphic phantom, direct MD resulted in incorrect interpretation of contrast and soft tissue, while SSQI successfully distinguished them and reduced RMSE in material-specific images by 38% to 92%. For the flow phantom, SSQI was able to perform accurate dynamic quantitative imaging, separating contrast from the background. Conclusions: We demonstrated the potential of SSQI for robust quantitative x-ray imaging. The simplicity of SSQI may enable its widespread adoption, including radiography and dynamic imaging such as real-time image guidance and cone-beam CT.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑