用光谱CT数据重建精确的综合Hounsfield单位
Reconstructing Accurate Synthetic Hounsfield Units with Spectral CT Data
摘要:
目的:旨在评价一种建议的方法,该方法用于重建CT数,该CT数准确地模拟来自光谱CT数据的传统CT数,如传统系统在没有射束硬化效应的情况下所产生的那样。方法:我们实现了模拟的理想和非理想的光子计数多仓系统,后者的基础上的光子计数硅探测器,并比较它们与模拟的理想的传统能量积分系统与杯形校正算法的方法。我们比较系统使用的数学模型的大小和组成,产生严重的杯形和射束硬化伪影时,成像与传统的系统没有杯形校正。结果:所得到的图像显示的CT值对于不同范围的组织始终准确,并且没有射束硬化伪影。结论:这种方法可以促进在从传统CT到光谱CT的过渡期间使用关于各种器官和条件的绝对CT数的既定经验法则。
英文摘要:
Purpose: To evaluate a proposed method to reconstruct CT numbers that accurately mimic conventional CT numbers from spectral CT data, as would have been produced by a conventional system without effects of beam hardening. Approach: We implement the proposed method for simulated ideal and non-ideal photon counting multi-bin systems, the latter based on a photon counting Silicon detector, and compare them with a simulated ideal conventional energy integrating system with a cupping correction algorithm. We compare the systems using a mathematical phantom of a size and composition that produces severe cupping and beam hardening artefacts when imaged with a conventional system with no cupping correction. Results: The resulting images show CT numbers that are consistently accurate for a varying range of tissues and are free of beam hardening artefacts. Conclusions: This method could facilitate use of established rules-of-thumb regarding absolute CT numbers for various organs and conditions during the transition from conventional CT to spectral CT.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。