电子显微镜中的复杂观测 VIII:最大化相位衬度灵敏度的新型 Hilbert 相位板

Complex Observation in Electron Microscopy VIII: Novel Hilbert Phase-plates to Maximize Phase-contrast Sensitivity

Kuniaki Nagayama

AI总结:

针对相位板透射电镜灵敏度争议,提出结合对称Hilbert相位板实验的新方法,通过理论与模拟对比证明其在π/2附近优于Scherzer欠焦,显著提升了相位衬度灵敏度。

摘要:

自从有报告指出 Volta 相位板相位衬度的灵敏度低于无相位板相位衬度,从而导致回归传统欠焦相位衬度以来,相位板透射电子显微镜最近出现了倒退。那么 Hilbert 相位板相位衬度又如何呢?我们报告称,如果将使用一对对称 Hilbert 相位板(其相位设置为小于 π 的值)进行的两次实验结合起来,Hilbert 相位板方法得以存续。在灵敏度方面,对分别代表 Volta 相位板和 Scherzer 欠焦的使用对称 Hilbert 相位板和 Zernike 相位板的三种相位衬度方法进行了理论比较,理论依据是衬度传递理论,计算比较则在专为相位板透射电子显微镜设计的模拟器上进行。对于这两种相位板相位衬度,通过改变相位板相位来搜索给出最高灵敏度的相位。结果发现,对称 Hilbert 相位板相位衬度在 π/2 附近的相位中优于 Scherzer 欠焦相位衬度。另一方面,发现 Zernike 相位板相位衬度在从 0 到 π 的整个相位范围内明显劣于 Scherzer 欠焦相位衬度。此外,将新型 Hilbert 相位板方法与同样需要两次实验的复杂观测透射电子显微镜进行了比较,并研究了对称 Hilbert 相位板相位衬度具有更高灵敏度的起源。

英文摘要:

Phase-plate transmission electron microscopy has recently regressed since a report that Volta phase-plate phase-contrast is less sensitive than non-phase-plate phase contrast, which leads to conventional defocusing phase-contrast. What about Hilbert phase-plate phase-contrast? We report that the Hilbert phase-plate method can survive if two experiments using a pair of symmetric Hilbert phase-plates, of which phase is set to a value smaller than π, are combined. Three phase-contrast methods using the symmetric Hilbert phase-plates and Zernike phase-plate representing Volta phase-plate and Scherzer defocus respectively were compared in sensitivity theoretically relying on a contrast transfer theory and computationally on a simulator specifically designed for phase plate transmission electron microscopy. For the two phase-plate phase-contrasts, the phase that gives the highest sensitivity was searched for by changing the phase-plate phase. As a result, the symmetric Hilbert phase-plates phase-contrast was found to outperform Scherzer defocus phase-contrast in the phase around π/2. On the other hand, Zernike phase-plate phase-contrast was found considerably inferior to Scherzer defocus phase-contrast in the entire phase range from 0 to π. Furthermore, the novel Hilbert phase-plate method was compared with complex observation transmission electron microscopy, which also requires two experiments, and the origin of the higher sensitivity of symmetric Hilbert phase-plates phase-contrast was examined.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑