便携式氧传感装置用于改善昏迷综合征和其他缺氧相关疾病的评估

Portable Oxygen-Sensing Device for the Improved Assessment of Compartment Syndrome and other Hypoxia-Related Conditions

Lilian Witthauer, Juan Pedro Cascales, Emmanuel Roussakis, Xiaolei Li, Avery Goss, Yenyu Chen, Conor L. Evans

摘要:

肌内氧的测量在急性筋膜室综合征的早期诊断中起关键作用,急性筋膜室综合征是严重创伤后的常见疾病,导致缺血和长期后果,包括横纹肌溶解、肢体丧失和死亡。然而,迄今为止,还没有现有的氧传感器被批准用于这种目的。为了解决需要改善评估筋膜室综合征,便携式光纤装置肌内血氧测量的发展。该装置是基于磷光猝灭,其中光纤的尖端涂有聚(甲基丙烯酸丙酯)(PPMA)矩阵含有明亮的发射Pt(II)-核心卟啉。光电电路是高度便携式的,基于微型光谱仪和微控制器,通过智能手机读出。体内止血带猪模型的结果表明,传感器在0-880 mmHg的生理氧分压范围内敏感,并对肌内氧变化表现出适当且可重现的响应。基于寿命测量的商业实验室氧传感器没有如预期的那样响应。

英文摘要:

Measurement of intramuscular oxygen could play a key role in the early diagnosis of acute compartment syndrome, a common condition occurring after severe trauma leading to ischemia and long-term consequences including rhabdomyolysis, limb loss, and death. However, to date, there is no existing oxygen sensor approved for such a purpose. To address the need to improve the assessment of compartment syndrome, a portable fiber-optic device for intramuscular oxygen measurements was developed. The device is based on phosphorescence quenching, where the tip of an optical fiber was coated with a poly(propyl methacrylate) (PPMA) matrix containing a brightly emitting Pt(II)-core porphyrin. The optoelectronic circuit is highly portable and is based on a microspectrometer and a microcontroller readout with a smartphone. Results from an in vivo tourniquet porcine model show that the sensor is sensitive across the physiological oxygen partial pressure range of 0-880 mmHg and exhibits an appropriate and reproducible response to changes in intramuscular oxygen. A commercial laboratory oxygen sensor based on a lifetime measurement did not respond as expected.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑