用于前列腺的准实时多频三维剪切波绝对振动弹性成像(S-WAVE)系统
Quasi-Real Time Multi-Frequency 3D Shear Wave Absolute Vibro-Elastography (S-WAVE) System for Prostate
AI总结:
本文提出一种用于前列腺的准实时多频三维剪切波绝对振动弹性成像(S-WAVE)系统,通过经会阴机械振动器产生多频剪切波,结合带通采样和斑点追踪算法实现组织弹性定量测量,临床验证表明其在前列腺癌识别上具有较高的诊断准确性。
摘要:
本文描述了一种新型准实时系统,用于定量和体积测量前列腺组织弹性。组织弹性通过使用局部频率估计器测量前列腺腺体内稳态剪切波的三维局部波长来计算。该剪切波由机械音圈振动器经会阴传输多频振动产生。射频数据直接从BK Medical 8848经直肠超声换能器流式传输至外部计算机,利用斑点追踪算法测量由激励引起的组织位移。系统采用带通采样,消除了对超快帧速率以追踪组织运动的需求,并允许在低于Nyquist率的采样频率下进行精确重建。由计算机控制的旋转电机用于旋转换能器的矢状阵列并获取三维数据。使用两个CIRS仿体验证了弹性测量的准确性以及将该系统用于体内前列腺成像的功能可行性。该系统已在两项独立的临床研究中作为癌症识别方法使用。本文呈现的结果表明,我们的刚度测量值与病理结果确定的癌症可能性之间存在数值和视觉相关性。使用该系统的初步已发表结果显示,在外周带前列腺癌识别方面,受试者工作特征曲线下面积为0.82+/-0.01。
英文摘要:
This article describes a novel quasi-real time system for quantitative and volumetric measurement of tissue elasticity in the prostate. Tissue elasticity is computed by using a local frequency estimator to measure the three dimensional local wavelengths of a steady-state shear wave within the prostate gland. The shear wave is created using a mechanical voice coil shaker which transmits multi-frequency vibrations transperineally. Radio frequency data is streamed directly from a BK Medical 8848 trans-rectal ultrasound transducer to an external computer where tissue displacement due to the excitation is measured using a speckle tracking algorithm. Bandpass sampling is used that eliminates the need for an ultra fast frame rate to track the tissue motion and allows for accurate reconstruction at a sampling frequency that is below the Nyquist rate. A roll motor with computer control is used to rotate the sagittal array of the transducer and obtain the 3D data. Two CIRS phantoms were used to validate both the accuracy of the elasticity measurement as well as the functional feasibility of using the system for in vivo prostate imaging. The system has been used in two separate clinical studies as a method for cancer identification. The results, presented here, show numerical and visual correlations between our stiffness measurements and cancer likelihood as determined from pathology results. Initial published results using this system include an area under the receiver operating characteristic curve of 0.82+/-0.01 with regards to prostate cancer identification in the peripheral zone.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。