两种多路复用便携式冷原子真空标准的比较

Comparison of two multiplexed portable cold-atom vacuum standards

Lucas H. Ehinger, Bishnu P. Acharya, Daniel S. Barker, James A. Fedchak, Julia Scherschligt, Eite Tiesinga, Stephen Eckel

摘要:

比较了两种基于超冷$^7$ Li原子的便携式冷原子真空标准(pCAVS)的真空测量结果。pCAVS是基于量子的标准,使用先验散射计算将测量的冷原子从保守阱的损失率转换为背景气体压力。我们的pCAVS设备共享相同的激光系统并同时测量真空。两个pCAVS一起检测到泄漏,泄漏速率约为$10^{-6}$ Pa L/s。修复泄漏后,pCAVS测量的压力约为40 nPa,不确定度为2.6%。两个pCAVS在其不确定性范围内一致,即使在交换它们的一些组成部分时也是如此。发现pCAVS的操作会导致一些额外的放气,大约为$10^{-8}$ Pa L/s,这可以在将来通过更好的热管理来减轻。

英文摘要:

We compare the vacuum measured by two portable cold-atom vacuum standards (pCAVS) based on ultracold $^7$Li atoms. The pCAVS are quantum-based standards that use a priori scattering calculations to convert a measured loss rate of cold atoms from a conservative trap into a background gas pressure. Our pCAVS devices share the same laser system and measure the vacuum concurrently. The two pCAVS together detected a leak with a rate on the order of $10^{-6}$ Pa L/s. After fixing the leak, the pCAVS measured a pressure of about 40 nPa with 2.6 % uncertainty. The two pCAVS agree within their uncertainties, even when swapping some of their component parts. Operation of the pCAVS was found to cause some additional outgassing, on the order of $10^{-8}$ Pa L/s, which can be mitigated in the future by better thermal management.