arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2202.12879eess.SYcs.SY
收藏

1 kHz视网膜激光治疗的模型预测控制

Model predictive control for retinal laser treatment at 1 kHz

Manuel Schaller, Viktoria Kleyman, Mario Mordmüller, Christian Schmidt, Mitsuru Wilson, Ralf Brinkmann, Matthias A. Müller, Karl Worthmann

摘要:

激光光凝是一种用于治疗视网膜疾病的技术。虽然这通常是手动完成或使用简单的控制方案,我们追求基于优化的方法,即模型预测控制(MPC),以强制限制峰值温度,因此,在医疗过程中的安全性-尽管组织的点依赖性吸收。为此,需要1 kHz的重复率,这使得实时要求成为一个主要挑战。我们提出了一个量身定制的MPC方案,使用参数模型简化,扩展卡尔曼滤波器的参数和状态估计,并适当地构造阶段成本,并验证其适用性在模拟和实验与猪眼。此外,我们给出了一些见解,专门为快速数值计算的实施。

英文摘要:

Laser photocoagulation is a technique applied in the treatment of retinal diseases. While this is often done manually or using simple control schemes, we pursue an optimization-based approach, namely Model Predictive Control (MPC), to enforce bounds on the peak temperature and, thus, safety during the medical treatment procedure - despite the spot-dependent absorption of the tissue. To this end, a repetition rate of 1 kHz is desirable rendering the real-time requirements a major challenge. We present a tailored MPC scheme using parametric model reduction, an extended Kalman filter for the parameter and state estimation, and suitably constructed stage costs and verify its applicability both in simulation and experiments with porcine eyes. Moreover, we give some insight on the implementation specifically tailored for fast numerical computations.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑