用于生理传感的低阻抗3D干电极的快速制造
Rapid Fabrication of Low Impedance, 3D Dry Electrodes for Physiological Sensing
摘要:
可以研究身体和诊断疾病的医疗电生理传感器依赖于持续低阻抗的电极-皮肤界面。临床标准湿电极使用水凝胶和皮肤磨损来改善界面,从而改善记录的信号质量。这些电极对自我管理具有挑战性,并妨碍家庭护理。可穿戴干电极更实用;然而,它们显示出比湿电极更高的阻抗,并且定制成本高。这项工作提出了一种制造方法,用于快速生产低阻抗,解剖学上适合干电极,不需要水凝胶。通过使用3D打印的化学镀铜和镀金,可以以现有真空沉积技术的一小部分成本为个人优化生物相容性电极。在典型情况下,与临床标准设备一起评估用该过程制成的示例3D干电极,以比较电气性能和舒适度。所得的干电极在50 Hz下表现出66.7k Ω的平均电极-皮肤阻抗,并且在没有任何水凝胶的情况下DC偏移为-20mV,当面积归一化时,其在没有皮肤磨损的湿电极所实现的范围内。
英文摘要:
Medical electrophysiological sensors that can study the body and diagnose diseases depend on consistently low impedance electrode-skin interfaces. Clinical-standard wet electrodes use hydrogels and skin abrasion to improve the interface and thus the recorded signal quality. These electrodes are challenging to self-administer and impede in-home care. Wearable dry electrodes are more practical; however, they show higher impedances than wet electrodes and are costly to customize. This work presents a fabrication method for rapidly producing low impedance, anatomically fit dry electrodes that do not require hydrogels. By using electroless copper and gold plating with 3D printing, biocompatible electrodes can be optimized for individuals at a fraction of the cost of existing vacuum deposition-based techniques. Example 3D dry electrodes made with this process are evaluated alongside clinical-standard devices in typical scenarios to compare electrical performance and comfort. The resulting dry electrodes exhibited an average electrode-skin impedance of 66.7 kΩ at 50Hz and DC offset of -20 mV without any hydrogel, which, when area normalized, is within the range achieved by wet electrodes without skin abrasion.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。