用于小分子鉴定工作流程的碰撞截面特异性
Collision cross section specificity for small molecule identification workflows
摘要:
分子碰撞截面(CCS)的物理化学性质越来越多地用于辅助小分子鉴定;然而,关于其在促进此类鉴定方面的真正效用的程度仍然存在问题,特别是考虑到其与质量的相关性。为了研究CCS对给定文库内的独特性的贡献,我们使用多方向网格搜索测量了其作为CCS值(来自测量或预测)、CCS方差、母体质量、质量误差和/或参考数据库大小中的误差的函数的辨别能力。虽然实验CCS数据库存在,但它们目前很小;因此,我们使用CCS预测工具DarkChem提供理论CCS值用于本研究。然后修改这些预测的CCS值以反映实验方差。通过在各种精度下用CCS单独基于质量的库内增强我们的搜索,我们发现,(i)使用多个加合物(即相同母体化合物的替代电离形式),与使用单一加合物相比,极大地提高了特异性,以及(ii)当可以实现低CCS误差(例如1%的复合误差)时,甚至单个CCS也导致显著的特异性增强。基于这些结果,我们建议使用多个加合物来建立存在的证据,因为每个加合物提供每个维度的额外信息。此外,无论CCS是否被视为识别指标,仍应考虑离子迁移谱法与质谱法结合使用时的效用,这是由于其优点,例如增加的峰分辨率、灵敏度(例如,在任何给定时间减少检测器上的负载)、数据独立MS/MS光谱采集的改进以及更清洁的串联质谱裂解模式。
英文摘要:
The physical-chemical property of molecular collision cross section (CCS) is increasingly used to assist in small molecule identification; however, questions remain regarding the extent of its true utility in contributing to such identifications, especially given its correlation with mass. To investigate the contribution of CCS to uniqueness within a given library, we measured its discriminatory capacity as a function of error in CCS values (from measurement or prediction), CCS variance, parent mass, mass error, and/or reference database size using a multi-directional grid search. While experimental CCS databases exist, they are currently small; thus, we used a CCS prediction tool, DarkChem, to provide theoretical CCS values for use in this study. These predicted CCS values were then modified to mirror experimental variance. By augmenting our search within a library based on mass alone with CCS at a variety of accuracies, we found that, (i) the use of multiple adducts (i.e. alternative ionized forms of the same parent compound) for the same molecule, compared to using a single adduct, greatly improves specificity and (ii) even a single CCS leads to a significant specificity boost when low CCS error (e.g. 1% composite error) can be achieved. Based on these results, we recommend using multiple adducts to build up evidence of presence, as each adduct supplies additional information per dimension. Additionally, the utility of ion mobility spectrometry when coupled with mass spectrometry should still be considered, regardless of whether CCS is considered as an identification metric, due to advantages such as increased peak resolution, sensitivity (e.g. from reducing load on the detector at any given time), improvements in data-independent MS/MS spectra acquisition, and cleaner tandem mass spectral fragmentation patterns.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。