arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2109.11651cond-mat.softphysics.data-anphysics.ins-det
收藏

光子计数散粒噪声对Total Internal Reflection显微术的影响

Effect of Photon Counting Shot Noise on Total Internal Reflection Microscopy

Fan Cui, David J. Pine

摘要:

全内反射显微镜(TIRM)测量胶体颗粒和透明基底之间的距离变化,其通过测量当颗粒被由基底界面处全内反射的光产生的倏逝场照射时由颗粒散射的光的强度来进行。根据这些测量,可以测量胶体颗粒和基板之间的高度依赖性有效电势$φ(z)$。TIRM能够分辨高度$z$和有效电势$φ(z)$的空间分辨率受到用于测量散射光强度的光子计数过程的固有散粒噪声散粒噪声的限制。我们开发了一个模型来确定TIRM可以测量$φ(z)$的空间分辨率,并通过仿真和实验验证其有效性。我们进一步建立光子计数统计和强度积分时间$τ$在TIRM测量中的关键作用,这是缩小光子计数分布的宽度和捕获探针粒子的瞬时位置之间的权衡。

英文摘要:

Total internal reflection microscopy (TIRM) measures changes in the distance between a colloidal particle and a transparent substrate by measuring the intensity of light scattered by the particle when it is illuminated by the evanescent field that is created from light totally internally reflected at the substrate interface. From these measurements, the height-dependent effective potential $φ(z)$ between the colloidal particle and the substrate can be measured. The spatial resolution with which TIRM can resolve the height $z$ and effective potential $φ(z)$ is limited by the intrinsic shot noise of the photon counting process used to measure the scattered light intensity. We develop a model to determine the spatial resolution with which TIRM can measure $φ(z)$ and verify its validity with simulations and experiments. We further establish the critical role of photon-counting statistics and the intensity integration time $τ$ in TIRM measurements, which is a trade-off between narrowing the width of the photon counting distribution and capturing the instantaneous position of the probe particle.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑