在头部CT扫描中使用基于网络的Waza-AriV2软件估计器官吸收剂量
Estimation of absorbed dose in organs using Waza-AriV2 web-based software in head CT scans
摘要:
最近,Yamashita等人报告了在最频繁的临床CT扫描方案中暴露的尸检中(使用OSL剂量测定法)直接测量单个器官的电离辐射剂量,这些器官对应于全身、头部、胸部和腹部。作者将实验结果与Waza-AriV 2软件中计算的值进行了比较,Waza-AriV 2软件是一种基于蒙特卡罗方法的系统,可根据CT设备中使用的暴露参数估计器官剂量。然而,每次检查中计算的吸收剂量结果与死后直接测量的剂量有显著差异。在本通信中,我们在正确解释的情况下对其中一个采集参数(mA)进行了计算,这导致Yamashita等人报告的估计值和测量值之间存在显着差异。我们还对Waza-AriV 2和NCICT软件进行了比较。
英文摘要:
Recently, Yamashita et al. reported direct measurements of ionizing radiation doses to individual organs in postmortem (using OSL dosimetry) that were exposed in the most frequent clinical CT scan protocols, corresponding to the whole body, head, thorax and abdomen. The authors compare the experimental results with the values calculated in the Waza-AriV2 software, which is a system based on the MonteCarlo method that performs the estimation of organ doses from the exposure parameters used in the CT equipment. However, the results of the absorbed doses calculated in each examination had significant variations with respect to the doses measured directly in postmortem. In this communication, we made a calculation under the correct interpretation in one of the acquisition parameters (mA) which leads to a significant discrepancy between the estimated and measured values reported by Yamashita et al. We also make the comparison between the Waza-AriV2 and NCICT software.