早产儿脑氧合信号去饱和瞬态信号的稀疏去噪方法
Sparse-Denoising Methods for Extracting Desaturation Transients in Cerebral Oxygenation Signals of Preterm Infants
摘要:
早产儿在生命的最初几天内由于脑氧输送不良而具有发生脑损伤的高风险。近红外光谱(NIRS)是一种用于监测局部组织氧合的成熟技术。脑近红外光谱信号的详细波形分析可以提高该方法在准确预测脑损伤中的临床应用价值。在极早产儿中经常观察到短暂性脑氧饱和度下降,但其临床意义尚不清楚。本研究的目的是检查和比较两种不同的方法在隔离和提取瞬态偏转近红外光谱信号的性能。我们优化了三种不同的同时低通滤波和全变差去噪(LPF_TVD)方法,并将其性能与最近提出的一种使用奇异谱分析和离散余弦变换(SSA_DCT)的方法进行了比较。LPF_TVD方法的参数使用合成的类NIRS信号通过网格搜索进行优化。SSA_DCT方法进行了修改后的处理过程中,以增加稀疏提取的成分。我们的分析使用合成的NIRS类数据集,结果显示LPF_TVD方法优于改良的SSA_DCT方法:LPF_TVD方法的中位均方误差为0.97(95% CI:0.86至1.07),低于改良的SSA_DCT方法的中位均方误差1.48(95% CI:1.33至1.63),P<0.001。双低通滤波器和全变分去噪方法的计算效率比SSA_DCT方法高出3到4个数量级。需要更多的研究来检查这些方法在提取真实NIRS信号中的氧去饱和度的有效性。
英文摘要:
Preterm infants are at high risk of developing brain injury in the first days of life as a consequence of poor cerebral oxygen delivery. Near-infrared spectroscopy (NIRS) is an established technology developed to monitor regional tissue oxygenation. Detailed waveform analysis of the cerebral NIRS signal could improve the clinical utility of this method in accurately predicting brain injury. Frequent transient cerebral oxygen desaturations are commonly observed in extremely preterm infants, yet their clinical significance remains unclear. The aim of this study was to examine and compare the performance of two distinct approaches in isolating and extracting transient deflections within NIRS signals. We optimized three different simultaneous low-pass filtering and total variation denoising (LPF_TVD) methods and compared their performance with a recently proposed method that uses singular-spectrum analysis and the discrete cosine transform (SSA_DCT). Parameters for the LPF_TVD methods were optimized over a grid search using synthetic NIRS-like signals. The SSA_DCT method was modified with a post-processing procedure to increase sparsity in the extracted components. Our analysis, using a synthetic NIRS-like dataset, showed that a LPF_TVD method outperformed the modified SSA_DCT method: median mean-squared error of 0.97 (95% CI: 0.86 to 1.07) was lower for the LPF_TVD method compared to the modified SSA_DCT method of 1.48 (95% CI: 1.33 to 1.63), P<0.001. The dual low-pass filter and total variation denoising methods are considerably more computational efficient, by 3 to 4 orders of magnitude, than the SSA_DCT method. More research is needed to examine the efficacy of these methods in extracting oxygen desaturation in real NIRS signals.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。