ReBeatICG:实时低复杂度逐拍阻抗图描绘算法
ReBeatICG: Real-time Low-Complexity Beat-to-beat Impedance Cardiogram Delineation Algorithm
摘要:
这项工作提出了ReBeatICG,一个实时的,低复杂度的心跳到心跳阻抗心动图(ICG)描绘算法,允许血液动力学参数监测。与文献中发现的依赖于同步心电图信号(ECG)记录的大多数算法相比,所提出的程序仅依赖于ICG信号。ReBeatICG的设计考虑到了在超低功耗微控制器(MCU)上的实现。所开发的算法的检测精度进行测试,对心脏病专家手动标记的点。它对B、C、X和O点的检测Gmean准确率分别为94.9%、98.6%、90.3%和84.3%。此外,基于注释的特征点计算几个血流动力学参数,并与从心脏病学家的注释生成的值进行比较。ReBeatICG对HR、LVET、IVRT和相对C点振幅的平均误差率分别为0.11 ms、9.72 ms、8.32 ms和3.97%。
英文摘要:
This work presents ReBeatICG, a real-time, low-complexity beat-to-beat impedance cardiography (ICG) delineation algorithm that allows hemodynamic parameters monitoring. The proposed procedure relies only on the ICG signal compared to most algorithms found in the literature that rely on synchronous electrocardiogram signal (ECG) recordings. ReBeatICG was designed with implementation on an ultra-low-power microcontroller (MCU) in mind. The detection accuracy of the developed algorithm is tested against points manually labeled by cardiologists. It achieves a detection Gmean accuracy of 94.9%, 98.6%, 90.3%, and 84.3% for the B, C, X, and O points, respectively. Furthermore, several hemodynamic parameters were calculated based on annotated characteristic points and compared with values generated from the cardiologists' annotations. ReBeatICG achieved mean error rates of 0.11ms, 9.72ms, 8.32ms, and 3.97% for HR, LVET, IVRT, and relative C-point amplitude, respectively.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。