人脑中基于FLASH的MT饱和度对3 T下B1-inhomogeneity剩余偏倚的校正
Correction of FLASH-based MT saturation in human brain for residual bias of B1-inhomogeneity at 3T
摘要:
背景资料:磁化转移(MT)饱和反映了由MT脉冲施加的MRI信号的额外饱和,并且主要由结合池的饱和驱动。MT脉冲对束缚极化的这种减少比吸收的微分B1平方定律预测的效率低。因此,B1不均匀性导致MT饱和度图中的残余偏差。我们推导出一个启发式校正,以减少这种偏见,广泛使用的多参数映射协议在3 T。方法:通过标称翻转角改变MT脉冲的幅度,以模拟B1的变化。MT饱和度对实际翻转角的依赖性具有线性校正项,该线性校正项针对灰质和白质分别确定。结果如下:MT饱和度与微分B1平方定律的偏差可以用MT脉冲的实际翻转角线性减小来描述。这种减少在灰质和白质之间没有显着差异。因此,事后校正不需要考虑不同的组织类型。偏置校正MT饱和度图似乎更对称,并突出了高度髓鞘化的管道。讨论:我们的校正涉及特定于MT脉冲的校准。虽然它也可以用于重新调整标称翻转角,但不同的MT脉冲和/或协议将需要单独校准。结论:建议的MT标测图B1校正可使用独立采集的翻转角标测图事后应用。
英文摘要:
Background: Magnetization transfer (MT) saturation reflects the additional saturation of the MRI signal imposed by an MT pulse and is largely driven by the saturation of the bound pool. This reduction of the bound polarization by the MT pulse is less efficient than predicted by the differential B1-square law of absorption. Thus, B1 inhomogeneities lead to a residual bias in the MT saturation maps. We derive a heuristic correction to reduce this bias for a widely used multi-parameter mapping protocol at 3T. Methods: The amplitude of the MT pulse was varied via the nominal flip angle to mimic variations in B1. The MT saturation's dependence on the actual flip angle features a linear correction term, which was determined separately for gray and white matter. Results: The deviation of MT saturation from differential B1-square law is well described by a linear decrease with the actual flip angle of the MT pulse. This decrease showed no significant differences between gray and white matter. Thus, the post hoc correction does not need to take different tissue types into account. Bias-corrected MT saturation maps appeared more symmetric and highlighted highly myelinated tracts. Discussion: Our correction involves a calibration that is specific for the MT pulse. While it can also be used to rescale nominal flip angles, different MT pulses and/or protocols will require individual calibration. Conclusion: The suggested B1 correction of the MT maps can be applied post hoc using an independently acquired flip angle map.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。