Eddy current补偿双扩散编码(DDE)MRI
Eddy current compensated double diffusion encoded (DDE) MRI
摘要:
目的:涡流可能导致扩散加权回波平面成像中的图像失真。提出了一种方法,以减少双扩散编码(DDE)的MRI实验和由此派生的微观分数各向异性(μFA)的影响。方法:采用两次重聚焦自旋回波法测量DDE。为了评估个体扩散编码对图像失真的影响,进行了水中塑料棒网格的测量。在6名健康志愿者的大脑中评估涡流补偿对μFA测量的影响。结果:涡流补偿的使用减少了信号变化。正如预期的那样,由第二编码引起的失真大于第一编码引起的失真,从而需要更强烈地补偿它们。然而,为了获得最佳结果,必须对两种编码进行补偿。伪影减少通过减少高估大大改善了心室和灰质中μFA的测量。未观察到补偿对白质中μFA绝对值的影响。结论:在DDE涡流测量中,建议对两种编码进行补偿。
英文摘要:
Purpose: Eddy currents might lead to image distortions in diffusion weighted echo planar imaging. A method is proposed to reduce their effects on double diffusion encoding (DDE) MRI experiments and the thereby derived microscopic fractional anisotropy (μFA) . Methods: The twice refocused spin echo scheme was adapted for DDE measurements. To assess the effect of individual diffusion encodings on the image distortions, measurements of a grid of plastic rods in water were performed. The effect of eddy current compensation on μFA measurements was evaluated in the brains of six healthy volunteers. Results: The use of an eddy current compensation reduced the signal variation. As expected, the distortions caused by the second encoding were larger than those of the first encoding entailing a stronger need to compensate for them. For an optimal result, however, both encodings had to be compensated. The artifact reduction strongly improved the measurement of the μFA in ventricles and grey matter by reducing the overestimation. An effect of the compensation on absolute μFA values in white matter was not observed. Conclusion: It is advisable to compensate both encodings in DDE measurements for eddy currents.