arXiv 2103.13691physics.app-ph
收藏

永磁体性能分布对移动式NMR磁体组件磁场均匀性的影响

Influence of the distribution of the properties of permanent magnets on the field homogeneity of magnet assemblies for mobile NMR

Yannick Klein, Leon Abelmann, Han Gardeniers

摘要:

我们优化了移动微流体核磁共振应用的两种典型设计的磁场均匀性:两个具有气隙的平行磁铁和改进的Halbach阵列。随着样本长度的影响,一般的设计准则将提出。为了进行公平的比较,样品的敏感长度已被选择为与磁体之间的间隙大小相同,以确保发射和接收单元以及基本电匀场组件有足够的空间。考虑到最终设备的紧凑性,定义了一个边长为间隙尺寸5倍的盒子,完整的磁体配置应该适合其中。在选定的边界条件下,简单的平行长方体配置在没有主动匀场的情况下达到最佳均匀性(0.5 $\mathrm{B_{s}}$,41 ppm),而伪Halbach配置具有最高的场强(0.9 $\mathrm{B_{s}}$,994 ppm),假设完美的磁体。然而,永磁体配置存在缺陷,例如磁化、制造和定位误差,这导致比使用固定优化参数集的模拟所预期的更差的磁场均匀性。我们提出了一个磁立方体的灵敏度分析和研究结果的磁化强度和磁化角度的变化,从不同的供应商购买的磁铁,由不同的材料和涂层,和不同的大小。我们进行了详细的蒙特卡罗模拟的磁性能的测量分布上提到的配置的效果。长方体设计显示平均均匀度为430 ppm(标准偏差)。350 ppm),伪Halbach的平均均一性为1086 ppm(标准偏差)。8 ppm)。

英文摘要:

We optimised the magnetic field homogeneity of two canonical designs for mobile microfluidic NMR applications: two parallel magnets with an air gap and a modified Halbach array. Along with the influence of the sample length, general design guidelines will be presented. For a fair comparison the sensitive length of the sample has been chosen to be the same as the gap size between the magnets to ensure enough space for the transmitting and receiving unit, as well as basic electric shimming components. Keeping the compactness of the final device in mind, a box with an edge length 5 times the gap size has been defined, in which the complete magnet configuration should fit. With the chosen boundary conditions, the simple parallel cuboid configuration reaches the best homogeneity without active shimming (0.5$\mathrm{B_{s}}$, 41 ppm), while the Pseudo-Halbach configuration has the highest field strength (0.9$\mathrm{B_{s}}$, 994 ppm), assuming perfect magnets. However, permanent magnet configurations suffer from imperfections, such as magnetisation, fabrication and positioning errors, which results in worse magnetic field homogeneities than expected from simulations using a fixed optimised parameter set. We present a sensitivity analysis for a magnetic cube and the results of studies of the variations in the magnetisation and angle of magnetisation of magnets purchased from different suppliers, composed of different materials and coatings, and of different sizes. We performed a detailed Monte Carlo simulation on the effect of the measured distribution of magnetic properties on the mentioned configurations. The cuboid design shows a mean homogeneity of 430 ppm (std dev. 350 ppm), the Pseudo-Halbach has a mean homogeneity of 1086 ppm (std dev. 8 ppm).

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑