峰值拾取多电压阈值数字化仪在辐射测量能量测定中的潜在优势
Potential Advantages of Peak Picking Multi-Voltage Threshold Digitizer in Energy Determination in Radiation Measurement
摘要:
多电压阈值(MVT)方法是一种利用一定的参考电压对信号进行采样的方法,已被广泛应用于临床前和临床数字正电子发射断层扫描(PET)系统。为了提高其能量测量性能,本文提出了一种峰值拾取MVT(PP-MVT)数字化仪。该方法首先通过简单的峰值采样电路在传统的MVT中增加一个采样峰值点(脉冲信号的最高点),该点携带幅度特征电压和幅度到达时间。其次,采用幅度偏差统计分析方法,比较不同重建模型的能量偏差,为不同幅度的信号脉冲选择自适应重建模型。用~(22)Na点源对示波器随机采样的3万个脉冲进行处理后,在450-650 KeV能量窗口内,该方法的能量分辨率达到17.50%,比相同阈值的传统MVT方法提高了2.44%;在相同的能量窗口中,我们得到的计数值为15225,而MVT的结果为14678。当PP-MVT涉及的阈值比传统的MVT少时,仍然可以保持更好的能量分辨率和更大的计数数的优点,这显示了PP-MVT数字化仪的鲁棒性和灵活性。该改进方法表明,添加特征峰信息可以提高信号采样和重构的性能,辐射测量中能量确定的更好性能可以证明这一点。
英文摘要:
The Multi-voltage Threshold (MVT) method, which samples the signal by certain reference voltages, has been well developed as being adopted in pre-clinical and clinical digital positron emission tomography(PET) system. To improve its energy measurement performance, we propose a Peak Picking MVT(PP-MVT) Digitizer in this paper. Firstly, a sampled Peak Point(the highest point in pulse signal), which carries the values of amplitude feature voltage and amplitude arriving time, is added to traditional MVT with a simple peak sampling circuit. Secondly, an amplitude deviation statistical analysis, which compares the energy deviation of various reconstruction models, is used to select adaptive reconstruction models for signal pulses with different amplitudes. After processing 30,000 randomly-chosen pulses sampled by the oscilloscope with a 22Na point source, our method achieves an energy resolution of 17.50% within a 450-650 KeV energy window, which is 2.44% better than the result of traditional MVT with same thresholds; and we get a count number at 15225 in the same energy window while the result of MVT is at 14678. When the PP-MVT involves less thresholds than traditional MVT, the advantages of better energy resolution and larger count number can still be maintained, which shows the robustness and the flexibility of PP-MVT Digitizer. This improved method indicates that adding feature peak information could improve the performance on signal sampling and reconstruction, which canbe proved by the better performance in energy determination in radiation measurement.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。