低能蒙特卡罗模拟中与光电效应有关的B型不确定性研究

A study of Type B uncertainties associated with the photoelectric effect in low-energy Monte Carlo simulations

Christian Valdes-Cortez, Iymad Mansour, Mark J. Rivard, Facundo Ballester, Ernesto Mainegra-Hing, Rowan M. Thomson, Javier Vijande

摘要:

本文的目的是估计低能光子截面(<200 keV)的当前最先进的蒙特卡罗代码中不同实现所产生的床剂量计算中的B型不确定性。蒙特卡罗模拟使用在低能量域中广泛使用的三种代码进行:PENMENTAL PE-2018、EGSnrc和MCNP。质量能量吸收系数的水,空气,石墨,和它们各自的比例;吸收剂量;和光子注量谱被认为是。使用类似的横截面进行了基准模拟。当考虑不同的横截面时,这些量之间观察到的差异被认为是相应的B型不确定性的良好估计量。估计吸收剂量(k=2)的保守B型不确定度为1.2%-1.7%(<50 keV)、0.6%-1.2%(50-100 keV)和0.3%(100-200 keV)。光子通量谱不存在值得考虑额外B型不确定性的临床相关差异,但能量低于25 keV的情况除外,其中获得的B型不确定性为0.5%。在30 keV以下,质量能量吸收系数显示B型不确定性(k=2),约为1.5%(水和空气)和2%(石墨),达到约1%(40-50 keV)和0.5%(50-75 keV)的值。B型不确定性的水石墨比观察到的能量低于30千电子伏,约0.7%(k=2)。与中间(约500 keV)或高(约1 MeV)的能量域,B型不确定性,由于不同的横截面的实施不能被认为是相对于A型不确定性,甚至其他来源的B型不确定性次主导。因此,这里报告的值应容纳在低能量光子剂量学研究的不确定性预算。

英文摘要:

The goal of this manuscript is to estimate Type B uncertainties in absorbed-dose calculations arising from the different implementations in current state-of-the-art Monte Carlo codes of low-energy photon cross-sections (<200 keV). Monte Carlo simulations are carried out using three codes widely used in the low-energy domain: PENELOPE-2018, EGSnrc, and MCNP. Mass energy-absorption coefficients for water, air, graphite, and their respective ratios; absorbed dose; and photon-fluence spectra are considered. Benchmark simulations using similar cross-sections have been performed. The differences observed between these quantities when different cross-sections are considered are taken to be a good estimator for the corresponding Type B uncertainties. A conservative Type B uncertainty for the absorbed dose (k=2) of 1.2%-1.7% (<50 keV), 0.6%-1.2% (50-100 keV), and 0.3% (100-200 keV) is estimated. The photon-fluence spectrum does not present clinically relevant differences that merit considering additional Type B uncertainties except for energies below 25 keV, where a Type B uncertainty of 0.5% is obtained. Below 30 keV, mass energy-absorption coefficients show Type B uncertainties (k=2) of about 1.5% (water and air), and 2% (graphite), reaching values about 1% (40-50 keV) and 0.5% (50-75 keV). Type B uncertainties for the water-to-graphite ratios are observed for energies below 30 keV, about 0.7% (k=2). In contrast with the intermediate (about 500 keV) or high (about 1 MeV) energy domains, Type B uncertainties due to the different cross-sections implementation cannot be considered subdominant with respect to Type A uncertainties or even to other sources of Type B uncertainties. Therefore, the values reported here should be accommodated within the uncertainty budget in low-energy photon dosimetry studies.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑