100 kW光功率静电天平测量机构的设计
Design of an electrostatic balance mechanism to measure optical power of 100 kW
摘要:
需要一种新的仪器,以适应在100 W以上的激光功率测量中增加便携性和准确性的需要。激光的反射和吸收提供了来自光子动量交换的可测量的力,该力与激光功率成正比,其可以用可追溯到SI的静电天平测量。我们的目标是$10^{-3}$与覆盖因子$k=2$的相对不确定性。为此目的,我们设计了一个整体的四杆机构,包括弹性圆形切口柔性铰链。该设计进行了优化,以解决驱动力测量不确定性的主要因素从平衡机制:角加载误差,平衡刚度,应力在柔性铰链,振动的敏感性,和敏感性的热梯度。弧形运动期间四杆机构自由端的寄生旋转受到加工公差的约束。分析模型表明,这对力测量的影响小于0.01%。倒立摆的摆动降低了系统的刚度,而不会过度增加倾斜灵敏度。弯曲的有限元建模用于确定使应力最小化的铰链取向,因此预期使滞后最小化。使用外部外壳来最小化温度梯度来减轻热效应,尽管没有对这种效应进行定量分析。这些分析表明,优化的机制,预计将有助于小于$10^{-3}$的相对不确定性,在最终的激光功率测量。
英文摘要:
A new instrument is required to accommodate the need for increased portability and accuracy in laser power measurement above 100 W. Reflection and absorption of laser light provide a measurable force from photon momentum exchange that is directly proportional to laser power, which can be measured with an electrostatic balance traceable to the SI. We aim for a relative uncertainty of $10^{-3}$ with coverage factor $k=2$. For this purpose, we have designed a monolithic parallelogram 4-bar linkage incorporating elastic circular notch flexure hinges. The design is optimized to address the main factors driving force measurement uncertainty from the balance mechanism: corner loading errors, balance stiffness, stress in the flexure hinges, sensitivity to vibration, and sensitivity to thermal gradients. Parasitic rotations in the free end of the 4-bar linkage during arcuate motion are constrained by machining tolerances. An analytical model shows this affects the force measurement less than 0.01 percent. Incorporating an inverted pendulum reduces the stiffness of the system without unduly increasing tilt sensitivity. Finite element modeling of the flexures is used to determine the hinge orientation that minimizes stress which is therefore expected to minimize hysteresis. Thermal effects are mitigated using an external enclosure to minimize temperature gradients, although a quantitative analysis of this effect is not carried out. These analyses show the optimized mechanism is expected to contribute less than $10^{-3}$ relative uncertainty in the final laser power measurement.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。