针对解剖噪声的乳腺X射线摄影优化
Optimization of mammography with respect to anatomical noise
摘要:
乳腺X射线摄影中的射束质量优化传统上考虑在均匀背景上检测被量子噪声遮蔽的目标。可以认为,该方案并不能很好地对应于临床成像任务,因为真实的乳房X线摄影图像包含解剖结构的复杂叠加,导致可能超过量子噪声的解剖噪声。使用新开发的光谱乳腺X射线摄影系统,我们测量了一组乳腺X射线摄影中解剖噪声的相关性和幅度。来自这些测量的结果被用作包括量子噪声以及解剖噪声的干扰模型优化的输入。我们发现,在这个框架内,肿瘤和微钙化的可检测性表现出非常不同的光束质量和剂量。小微钙化的结果与传统优化方法的结果相似,这是可以预期的,因为量子噪声在高空间频率下占主导地位。然而,对于较大的肿瘤,低频解剖噪声是限制因素。由于解剖结构具有与肿瘤对比度相似的能量依赖性,因此最佳X射线能量显著高于传统方法,并且有用能量区域比传统方法更宽。对组织体模的测量证实了这些理论结果。此外,由于量子噪声仅构成噪声的一小部分,因此可以在不牺牲肿瘤可检测性的情况下大幅降低剂量。因此,临床上使用的曝光设置不一定是该成像任务的最佳设置。然而,这些发现对整个乳腺X线摄影成像任务的影响在现阶段尚不清楚。
英文摘要:
Beam quality optimization in mammography traditionally considers detection of a target obscured by quantum noise on a homogenous background. It can be argued that this scheme does not correspond well to the clinical imaging task because real mammographic images contain a complex superposition of anatomical structures, resulting in anatomical noise that may dominate over quantum noise. Using a newly developed spectral mammography system, we measured the correlation and magnitude of the anatomical noise in a set of mammograms. The results from these measurements were used as input to an observer-model optimization that included quantum noise as well as anatomical noise. We found that, within this framework, the detectability of tumors and microcalcifications behaved very differently with respect to beam quality and dose. The results for small microcalcifications were similar to what traditional optimization methods would yield, which is to be expected since quantum noise dominates over anatomical noise at high spatial frequencies. For larger tumors, however, low-frequency anatomical noise was the limiting factor. Because anatomical structure has similar energy dependence as tumor contrast, optimal x-ray energy was significantly higher and the useful energy region wider than traditional methods suggest. Measurements on a tissue phantom confirmed these theoretical results. Furthermore, since quantum noise constitutes only a small fraction of the noise, the dose could be reduced substantially without sacrificing tumor detectability. Exposure settings used clinically are therefore not necessarily optimal for this imaging task. The impact of these findings on the mammographic imaging task as a whole is, however, at this stage unclear.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。