FRACAS强子治疗截面测量装置的系统研究

Geant4 Systematic Study of the FRACAS Apparatus for Hadrontherapy Cross Section Measurements

Edgar Barlerin, Samuel Salvador, Marc Labalme

摘要:

在这项工作中,我们报告的FRACAS装置,一个大的接受质谱仪,将被用来测量强子治疗的$^{12}$ C离子的碎裂截面的系统蒙特卡罗(MC)研究。放置在100 mbar反应室中的装置将由束流监视器、围绕磁体的跟踪器和飞行时间(TOF)壁制成。为了确定跟踪器所需的性能,使用了整个系统的Geant 4模拟和内部开发的算法。在保持束流监测器和TOF壁位置固定的情况下,提取了跟踪器位置和空间分辨率对轨迹重建和质量识别效率的影响。一个最佳的配置被发现,上游跟踪器应位于6厘米远离目标和间隔4厘米,而他们的空间分辨率应接近100 $μ$米。下游跟踪器的位置将不得不根据波束能量而改变,以保持高识别效率。它们的空间分辨率,即使与上游跟踪器相比不太重要,也应该在1 mm左右或更好。在这个最佳配置中,总的碎片识别效率超过90%,已获得从100至400 MeV/核子的束流能量范围内。

英文摘要:

In this work, we report on systematic Monte Carlo (MC) studies for the FRACAS apparatus, a large acceptance mass spectrometer that will be used to measure the fragmentation cross sections of $^{12}$C ions for hadrontherapy. The apparatus placed, in a 100 mbar reaction chamber, will be made of a beam monitor, trackers surrounding a magnet and a Time-Of-Flight (TOF) wall. In order to determine the required performances of the trackers, Geant4 simulations of the whole system and in-house developed algorithms were used. While keeping the beam monitor and TOF-wall positions fixed, the effects of the tracker positions and spatial resolutions on the trajectory reconstruction and mass identification efficiencies have been extracted. An optimal configuration was found where the upstream trackers should be located 6 cm away from the target and spaced 4 cm apart whereas their spatial resolutions should be close to 100 $μ$m. The positions of the downstream trackers will have to be changed according to the beam energy to preserve high identification efficiencies. Their spatial resolutions, even though of a lesser importance compared to the upstream trackers, should be around 1 mm or better. In this optimal configuration, an overall fragment identification efficiency above 90% has been obtained for beam energies ranging from 100 to 400 MeV/nucleon.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑