用于毫米波光谱辐射计大气分子多线观测的波导型多路复用器

Waveguide-Type Multiplexer for Multiline Observation of Atmospheric Molecules using Millimeter-Wave Spectroradiometer

Taku Nakajima, Kohei Haratani, Akira Mizuno, Kazuji Suzuki, Takafumi Kojima, Yoshinori Uzawa, Shin'ichiro Asayama, Issei Watanabe

摘要:

为了更好地了解中层大气臭氧丰度的变化机制,同时监测臭氧和其他与臭氧消耗有关的小分子物种是最基本和最关键的方法。为了扩大毫米波光谱辐射计的观测频率范围,实现多条分子谱线的同时观测,研制了一种波导型多路复用器。该多路复用器包含一个级联的四级边带分离滤波器电路。在电磁分析的基础上设计了波导电路,第1-4级的通频带为243-251 GHz、227-235 GHz、197-205 GHz和181-189 GHz。使用矢量网络分析仪测量了多路复用器的插入损耗和回波损耗,定义了每个观测频带,并适当指定了带宽。此外,在4K下测量了使用超导混频器的接收机噪声温度和镜像抑制比(IRR)。结果,由于多路复用器而导致的接收机噪声与仅混频器的接收机噪声相比的增加可以归因于多路复用器中的波导电路的传输损耗。在每个观察波段的中心,IRR均高于25 dB。这表明波导型多路复用器用于分离边带信号可以实现高且稳定的IRR性能。

英文摘要:

In order to better understand the variation mechanism of ozone abundance in the middle atmosphere, the simultaneous monitoring of ozone and other minor molecular species, which are related to ozone depletion, is the most fundamental and critical method. A waveguide-type multiplexer was developed for the expansion of the observation frequency range of a millimeter-wave spectroradiometer, for the simultaneous observation of multiple molecular spectral lines. The proposed multiplexer contains a cascaded four-stage sideband-separating filter circuit. The waveguide circuit was designed based on electromagnetic analysis, and the pass frequency bands of Stages 1-4 were 243-251 GHz, 227-235 GHz, 197-205 GHz, and 181-189 GHz. The insertion and return losses of the multiplexer were measured using vector network analyzers, each observation band was well-defined, and the bandwidths were appropriately specified. Moreover, the receiver noise temperature and the image rejection ratio (IRR) using the superconducting mixer at 4 K were measured. As a result, the increase in receiver noise due to the multiplexer compared with that of only the mixer can be attributed to the transmission loss of the waveguide circuit in the multiplexer. The IRRs were higher than 25 dB at the center of each observation band. This indicates that a high and stable IRR performance can be achieved by the waveguide-type multiplexer for the separation of sideband signals.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑