超极化$^{13}$ C MRI的区域团注跟踪和实时B $_1$校准方法

A Regional Bolus Tracking and Real-time B$_1$ Calibration Method for Hyperpolarized $^{13}$C MRI

Shuyu Tang, Eugene Milshteyn, Galen Reed, Jeremy Gordon, Robert Bok, Xucheng Zhu, Zihan Zhu, Daniel B. Vigneron, Peder E. Z. Larson

摘要:

目的:采集时间和B型$_1$校准是影响超极化$^{13}$ C MRI质量和准确性的两个关键因素。本项目的目标是开发一种新方法,使用区域推注跟踪触发Bloch-Siegert B $_1$标测和基于区域B $_1$测量的实时B $_1$校准,然后对体内超极化$^{13}C$代谢物进行动态成像。方法:所提出的方法实现了一个系统,允许实时数据处理和实时控制的顺序。还增加了推注到达时的实时中心频率校准。将所提出的框架应用于体内超极化$^{13}$ C成像的可行性在超极化[1- $^{13}$ C]丙酮酸盐注射后在临床3 T扫描仪上对健康大鼠、荷瘤小鼠和健康志愿者进行了测试。在人体研究中使用了多通道接收线圈。结果:该框架实现了基于区域团注峰值或团注到达的自动采集定时。通过实时中心频率校准,观察到实时重建图像中的模糊伪影减少。实时计算的B $_1$比例因子与实时采集的B $_1$图一致。使用B $_1$图谱进行翻转角校正,可更一致地定量代谢活性(即,乳酸盐-乳酸盐转化率,k $_{PL}$)。提供实验记录以演示实验期间的实时动作。结论:所提出的方法在动物和人类志愿者身上得到了成功的证明,并且预计将提高超极化信号的有效利用以及超极化$^{13}$ C成像的准确性和鲁棒性。

英文摘要:

Purpose: Acquisition timing and B$_1$ calibration are two key factors that affect the quality and accuracy of hyperpolarized $^{13}$C MRI. The goal of this project was to develop a new approach using regional bolus tracking to trigger Bloch-Siegert B$_1$ mapping and real-time B$_1$ calibration based on regional B$_1$ measurements, followed by dynamic imaging of hyperpolarized $^{13}C$ metabolites in vivo. Methods: The proposed approach was implemented on a system which allows real-time data processing and real-time control on the sequence. Real-time center frequency calibration upon the bolus arrival was also added. The feasibility of applying the proposed framework for in vivo hyperpolarized $^{13}$C imaging was tested on healthy rats, tumor-bearing mice and a healthy volunteer on a clinical 3T scanner following hyperpolarized [1-$^{13}$C]pyruvate injection. Multichannel receive coils were used in the human study. Results: Automatic acquisition timing based on either regional bolus peak or bolus arrival was achieved with the proposed framework. Reduced blurring artifacts in real-time reconstructed images were observed with real-time center frequency calibration. Real-time computed B$_1$ scaling factors agreed with real-time acquired B$_1$ maps. Flip angle correction using B$_1$ maps results in a more consistent quantification of metabolic activity (i.e, pyruvate-to-lactate conversion, k$_{PL}$). Experiment recordings are provided to demonstrate the real-time actions during the experiment. Conclusion: The proposed method was successfully demonstrated on animals and a human volunteer, and is anticipated to improve the efficient use of the hyperpolarized signal as well as the accuracy and robustness of hyperpolarized $^{13}$C imaging.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑