arXiv 2007.11527physics.med-ph
收藏

中心静脉压变化时颈静脉血流动力学的光学血流动力学成像

Optical Hemodynamic Imaging of Jugular Venous Dynamics During Altered Central Venous Pressure

Robert Amelard, Andrew D Robertson, Courtney A Patterson, Hannah Heigold, Essi Saarikoski, Richard L Hughson

摘要:

提出了一种用于定量评估颈静脉对中心静脉压改变的反应的光学成像系统。所提出的系统评估从颈静脉波形与空间校准程序来归一化入射组织照明的次表面光学吸收变化。右侧颈部的宽视野帧被捕获并使用新的柔性表面校准方法进行校准。推导了一个血流动力学光学模型,用于量化颈静脉光学衰减(JVA)信号,并生成空间颈静脉脉动图。在三种改变中心静脉压的心血管方案中评估JVA:急性中心血容量不足(下体负压)、静脉充血(头低位倾斜)和心脏充盈受损(Valsalva动作)。当与心电图时间对齐时,JVA波形表现出与颈静脉脉搏动力学一致的双相波特性。JVA与分级中心血容量不足(r=0.85,[0.72,0.95])、分级静脉充血(r=0.94,[0.84,0.99])和心脏充盈受损(r = 0.94,[0.85,0.99])期间的有创中心静脉压密切相关(中位数,四分位距)。分级急性血容量不足期间JVA降低与每搏输出量(SV)(r=0.85,[0.76,0.92])较基线降低(SV:79 $\pm$ 15 mL,JVA:0.56 $\pm$ 0.10 a.u.)至-40 mmHg抽吸(SV:59 $\pm$ 18 mL,JVA:0.47 $\pm$ 0.05 a.u.; p $<$ 0.01)。所提出的非接触式光学成像系统在三个改变中心静脉压的方案中证明了与有创中心静脉监测一致的颈静脉动力学。该系统在传统上需要导管插入术的临床相关条件下提供压力诱导颈静脉动力学的无创监测,从而能够在非手术环境中进行监测。

英文摘要:

An optical imaging system is proposed for quantitatively assessing jugular venous response to altered central venous pressure. The proposed system assesses sub-surface optical absorption changes from jugular venous waveforms with a spatial calibration procedure to normalize incident tissue illumination. Widefield frames of the right lateral neck were captured and calibrated using a novel flexible surface calibration method. A hemodynamic optical model was derived to quantify jugular venous optical attenuation (JVA) signals, and generate a spatial jugular venous pulsatility map. JVA was assessed in three cardiovascular protocols that altered central venous pressure: acute central hypovolemia (lower body negative pressure), venous congestion (head-down tilt), and impaired cardiac filling (Valsalva maneuver). JVA waveforms exhibited biphasic wave properties consistent with jugular venous pulse dynamics when time-aligned with an electrocardiogram. JVA correlated strongly (median, interquartile range) with invasive central venous pressure during graded central hypovolemia (r=0.85, [0.72, 0.95]), graded venous congestion (r=0.94, [0.84, 0.99]), and impaired cardiac filling (r=0.94, [0.85, 0.99]). Reduced JVA during graded acute hypovolemia was strongly correlated with reductions in stroke volume (SV) (r=0.85, [0.76, 0.92]) from baseline (SV: 79$\pm$15 mL, JVA: 0.56$\pm$0.10 a.u.) to -40 mmHg suction (SV: 59$\pm$18 mL, JVA: 0.47$\pm$0.05 a.u.; p$<$0.01). The proposed non-contact optical imaging system demonstrated jugular venous dynamics consistent with invasive central venous monitoring during three protocols that altered central venous pressure. This system provides non-invasive monitoring of pressure-induced jugular venous dynamics in clinically relevant conditions where catheterization is traditionally required, enabling monitoring in non-surgical environments.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑