外周动脉的机器人化超声成像--体模研究
Robotized Ultrasound Imaging of the Peripheral Arteries -- a Phantom Study
摘要:
对于患有外周动脉疾病的患者,诊断成像的首选是2D超声(US)。然而,对于适当的成像过程,需要熟练且有经验的超声医师。此外,这是一个高度依赖于用户的操作。自动扫描外周动脉的自动化US系统有可能克服这些限制。在这项工作中,我们扩展了以前提出的系统的分层图像分析管道的基础上卷积神经网络,以控制机器人。通过检查其在沿动脉扫描时将腿部体模的血管腔保持在US图像内的可行性,对该系统进行了评价。在扫描过程中获得的100%图像中,整个血管腔可见。当定义2.74 mm的不敏感边界时,血管中心与水平图像中心线之间的平均绝对距离在简单和复杂情况下分别为2.47 mm和3.90 mm。总之,该系统提供了使用无辐射方法进行全自动外周动脉成像的基础。
英文摘要:
The first choice in diagnostic imaging for patients suffering from peripheral arterial disease is 2D ultrasound (US). However, for a proper imaging process, a skilled and experienced sonographer is required. Additionally, it is a highly user-dependent operation. A robotized US system that autonomously scans the peripheral arteries has the potential to overcome these limitations. In this work, we extend a previously proposed system by a hierarchical image analysis pipeline based on convolutional neural networks in order to control the robot. The system was evaluated by checking its feasibility to keep the vessel lumen of a leg phantom within the US image while scanning along the artery. In 100 % of the images acquired during the scan process the whole vessel lumen was visible. While defining an insensitivity margin of 2.74 mm, the mean absolute distance between vessel center and the horizontal image center line was 2.47 mm and 3.90 mm for an easy and complex scenario, respectively. In conclusion, this system presents the basis for fully automatized peripheral artery imaging in humans using a radiation-free approach.