基于相位同步的多软件传感器无监督集成:一种用于心电图衍生呼吸信号的鲁棒方法
Unsupervised Ensembling of Multiple Software Sensors with Phase Synchronization: A Robust Approach For Electrocardiogram-derived Respiration
AI总结:
本研究提出一种基于相位同步的多EDR软件传感器无监督集成方法,通过质量筛选、图连接拉普拉斯相位对齐与融合生成高质量EDR信号,在两个大规模整夜多导睡眠图数据库上验证其性能优于现有算法,为从心电图提取鲁棒呼吸信息提供了实用方案。
摘要:
目的:本研究旨在融合不同心电图衍生呼吸(EDR)算法的输出,生成质量更高的单一EDR信号。方法:我们将每种EDR算法视为一个从不同观测角度记录呼吸活动的软件传感器,基于呼吸信号质量指数筛选出高质量软件传感器,采用基于图连接拉普拉斯算子的相位同步技术对质量最高的EDR信号进行对齐,最终融合这些经对齐的高质量EDR信号;我们将其输出称为同步集成EDR信号。研究在两个包含整夜多导睡眠图的大规模数据库上对所提算法进行评估,采用三种不同硬件传感器记录的呼吸信号评估算法性能,并将其与现有其他EDR算法对比。共开展5种场景的敏感性分析:对EDR信号取均值的融合场景,以及分别在无/有同步、无/有信号质量筛选条件下的4种EDR信号对齐场景。结果:在同步相关系数(γ-score)、最优传输(OT)距离和估计平均呼吸频率(EARR)得分三项指标上,同步集成EDR算法均优于现有EDR算法,且均具有统计学显著性。敏感性分析显示,信号质量筛选与EDR信号对齐对算法性能均至关重要,且均具有统计学显著性。结论:同步集成EDR可从心电图中提供鲁棒的呼吸信息。意义:相位同步不仅理论严谨,而且可用于设计鲁棒的EDR。
英文摘要:
Objective: We aimed to fuse the outputs of different electrocardiogram-derived respiration (EDR) algorithms to create one EDR signal that is of higher quality. Methods: We viewed each EDR algorithm as a software sensor that recorded breathing activity from a different vantage point, identified high-quality software sensors based on the respiratory signal quality index, aligned the highest-quality EDRs with a phase synchronization technique based on the graph connection Laplacian, and finally fused those aligned, high-quality EDRs. We refer to the output as the sync-ensembled EDR signal. The proposed algorithm was evaluated on two large-scale databases of whole-night polysomnograms. We evaluated the performance of the proposed algorithm using three respiratory signals recorded from different hardware sensors, and compared it with other existing EDR algorithms. A sensitivity analysis was carried out for a total of five cases: fusion by taking the mean of EDR signals, and the four cases of EDR signal alignment without and with synchronization and without and with signal quality selection. Results: The sync-ensembled EDR algorithm outperforms existing EDR algorithms when evaluated by the synchronized correlation (γ-score), optimal transport (OT) distance, and estimated average respiratory rate (EARR) score, all with statistical significance. The sensitivity analysis shows that the signal quality selection and EDR signal alignment are both critical for the performance, both with statistical significance. Conclusion: The sync-ensembled EDR provides robust respiratory information from electrocardiogram. Significance: Phase synchronization is not only theoretically rigorous but also practical to design a robust EDR.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。