用于适体药物监测的取样保持型CMOS电化学传感器的设计与分析

Design and Analysis of a Sample-and-Hold CMOS Electrochemical Sensor for Aptamer-based Therapeutic Drug Monitoring

Jun-Chau Chien, Sam W. Baker, H. Tom Soh, Amin Arbabian

摘要:

在本文中,我们提出了一个电化学电路的设计和分析,用于测量使用结构转换适配体的治疗药物的浓度。适体是单链核酸,其序列被选择为对分子靶标表现出高亲和力和特异性,并在结合时改变其构象。当与氧化还原报告子和电化学检测结合时,这种性质使得能够实现具有亚分钟时间分辨率的无试剂生物传感,用于体内治疗药物监测。具体而言,我们设计了一个计时电流法为基础的电化学电路,测量的直接变化的电子转移(ET)动力学的亚甲蓝报告共轭的适配体的远端。为了克服与大尺寸(> 0.25 mm 2)植入式电极接口时的高频噪声放大问题,我们提出了一种采样保持(S/H)电路技术,其中在记录氧化还原电流期间将所需的电极电位保持在无噪声电容器上。这允许断开反馈放大器以避免其噪声注入,同时降低总功耗。在65 nm CMOS中实现的原型电路演示了在2.5 kHz滤波带宽下15.2 pArms的单元电容不敏感输入参考噪声(IRN)电流。在人全血样品中进行测试,从记录的电流波形测量不同卡那霉素浓度下氧化还原标记的氨基糖苷类适体的ET动力学变化。通过采用主成分分析(PCA)来补偿采样误差,在0.22 mW的功耗下,1秒采集下的检测限(SNR = 1)为3.1 μ M。

英文摘要:

In this paper, we present the design and the analysis of an electrochemical circuit for measuring the concentrations of therapeutic drugs using structure-switching aptamers. Aptamers are single-stranded nucleic acids, whose sequence is selected to exhibit high affinity and specificity toward a molecular target, and change its conformation upon binding. This property, when coupled with a redox reporter and electrochemical detection, enables reagent-free biosensing with a sub-minute temporal resolution for in vivo therapeutic drug monitoring. Specifically, we design a chronoamperometry-based electrochemical circuit that measures the direct changes in the electron transfer (ET) kinetics of a methylene blue reporter conjugated at the distal-end of the aptamer. To overcome the high-frequency noise amplification issue when interfacing with a large-size (> 0.25 mm2) implantable electrode, we present a sample-and-hold (S/H) circuit technique in which the desired electrode potentials are held onto noiseless capacitors during the recording of the redox currents. This allows disconnecting the feedback amplifiers to avoid its noise injection while reducing the total power consumption. A prototype circuit implemented in 65-nm CMOS demonstrates a cell-capacitance-insensitive input-referred noise (IRN) current of 15.2 pArms at a 2.5-kHz filtering bandwidth. Tested in human whole blood samples, changes in the ET kinetics from the redox-labeled aminoglycoside aptamers at different kanamycin concentrations are measured from the recorded current waveforms. By employing principal component analysis (PCA) to compensate for the sampling errors, a detection limit (SNR = 1) of 3.1 uM under 1-sec acquisition is achieved at 0.22-mW power consumption.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑