ALICE μ子RPC的性能与老化研究

Performance and aging studies for the ALICE muon RPCs

Luca Quaglia, Antonio Bianchi, Alessandro Ferretti, Martino Gagliardi, Diego Stocco, Roberto Guida, Beatrice Mandelli, Laura Alvigini

AI总结:

本研究回顾了LHC Run2期间ALICE μ子RPC的性能稳定性,针对暗电流上升问题开展氩等离子体老化测试,为升级阶段的探测器更换决策提供依据。

摘要:

ALICE μ子触发(MTR)系统由72个阻性板室(RPC)探测器组成,分为两个探测站,每个探测站包含两个探测平面,每个平面有18个RPC。这些探测器以最大雪崩模式运行,使用的气体混合成分为89.7%的C₂H₂F₄、10%的异丁烷(i-C₄H₁₀)和0.3%的SF₆。在大型强子对撞机(LHC)的第二轮运行(Run2,2015-2018年)期间,研究人员持续监测了效率、暗电流等多项探测器性能指标。结果显示,探测效率保持了极佳的稳定性,但观测到探测器吸收的暗电流呈稳步上升趋势。\n自2018年底以来,LHC进入长停机阶段,在此期间ALICE实验将进行升级,以适配预计2021年启动的下一阶段数据采集工作。MTR正在进行前端电子学与读出电子学的重大升级,其功能也将发生转变,成为μ子鉴别器。本次升级仅计划更换受辐照最严重的RPC,因此开展专项研究以深入了解探测器状态至关重要。研究人员特别选取了两个RPC,用纯氩气长时间冲洗,并通过使气体完全电离产生等离子体。随后采用气相色谱-质谱联用仪分析输出气体,并借助离子选择电极工作站评估了等离子体与探测器内表面相互作用可能产生的氟化物的存在情况。\n本次报告将详细回顾LHC上ALICE μ子RPC的性能表现,同时也会讨论上述氩等离子体测试的流程与结果。

英文摘要:

The ALICE muon trigger (MTR) system consists of 72 Resistive Plate Chamber (RPC) detectors arranged in two stations, each composed of two planes with 18 RPCs per plane. The detectors are operated in maxi-avalanche mode using a mixture of 89.7% C$_2$H$_2$F$_4$, 10% i-C$_4$H$_{10}$ and 0.3% SF$_6$. A number of detector performance indicators, such as efficiency and dark current, have been monitored over time throughout the LHC Run2 (2015-18). While the efficiency showed very good stability, a steady increase in the absorbed dark current was observed. Since the end of 2018, the LHC has entered a phase of long shutdown, during which the ALICE experiment will be upgraded to cope with the next phase of data taking, expected in 2021. The MTR is undergoing a major upgrade of the front-end and readout electronics, and will change its functionalities, becoming a Muon Identifier. Only the replacement of the most irradiated RPCs is planned during the upgrade. It is therefore important to perform dedicated studies to gain further insights into the status of the detector. In particular, two RPCs were flushed with pure Ar gas for a prolonged period of time and a plasma was created by fully ionizing the gas. The output gas was analyzed using a Gas Chromatograph combined with a Mass Spectrometer and the possible presence of fluorinated compounds originating from the interaction of the plasma with the inner surfaces of the detector has been assessed using an Ion-Selective Electrode station. This contribution will include a detailed review of the ALICE muon RPC performance at the LHC. The procedure and results of the argon plasma test, described above, are also discussed.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑