Investigation of the Electrical Conduction Mechanisms in P-type Amorphous Germanium (a-Ge) Used as a-Ge Contacts for Ge Detectors

S. Bhattarai, R. Panth, W. -Z. Wei, H. Mei, D. -M. Mei, M. S. Raut, P. Acharya, W. -J. Wang

英文摘要:

Electrical conduction mechanisms in the disordered material system is experimentally studied for p-type amorphous germanium (a-Ge) used for high-purity Ge detector contacts. The localization length and the hopping parameters in a-Ge are determined using the surface leakage current measured from three high-purity planar Ge detectors. The temperature-dependent hopping distance and hopping energy are obtained for a-Ge fabricated as the electrical contact materials for high-purity Ge planar detectors. As a result, we find that the hopping energy in a-Ge increases as temperature increases while the hopping distance in a-Ge decreases as temperature increases. The localization length of a-Ge is on the order of $2.13^{-0.05}_{+0.07} A^\circ$ to $5.07^{-0.83}_{+2.58}A^\circ$, depending on the density of states near the Fermi energy level within bandgap. Using these parameters, we predict that the surface leakage current from a Ge detector with a-Ge contacts can be much smaller than one yocto amp (yA) at helium temperature, suitable for rare-event physics searches.