用于从临床血液样品中有效分离和分析循环肿瘤细胞的新型自动微滤装置的表征
Characterization of a novel automated microfiltration device for the efficient isolation and analysis of circulating tumor cells from clinical blood samples
摘要:
循环肿瘤细胞(CTC)的检测和分析可以实现广泛的癌症相关应用,包括在治疗期间鉴定获得性耐药性。然而,与大多数用于分离这些稀有细胞的技术相关的不可扩展的制造、延长的样品处理时间以及缺乏自动化,阻碍了它们向临床实践的过渡。这项工作描述了一种新型的基于膜的微滤装置,该装置由全自动样品处理单元和机器视觉成像系统组成,该系统允许从血液中有效分离和快速分析CTC。使用四种前列腺癌细胞系(包括PC-3、VCaP、DU-145和LNCaP)表征了该装置的性能,在处理来自健康供体的7.5 mL血液样品(掺入100个癌细胞)后,获得了高测定重现性和大于93%的捕获效率。由于手术过程中施加在细胞上的切应力最小,过滤后癌细胞仍然存活,而通过免疫染色识别癌细胞不受膜上捕获的非特异性事件数量的影响。我们还能够使用RT-PCR和Sanger测序从加标50个LNCaP细胞的7.5 mL血液样品中鉴定雄激素受体(AR)点突变T878 A。最后,在来自诊断患有转移性前列腺癌的患者的8个样品中的8个中检测到CTC(平均值$\pm$ SEM = 21 $\pm$ 2.957 CTC/mL,中值= 21 CTC/mL),从而验证该装置的潜在临床实用性。
英文摘要:
The detection and analysis of circulating tumor cells (CTCs) may enable a broad range of cancer-related applications, including the identification of acquired drug resistance during treatments. However, the non-scalable fabrication, prolonged sample processing times, and the lack of automation, associated with most of the technologies developed to isolate these rare cells, have impeded their transition into the clinical practice. This work describes a novel membrane-based microfiltration device comprised of a fully automated sample processing unit and a machine-vision-enabled imaging system that allows the efficient isolation and rapid analysis of CTCs from blood. The device performance was characterized using four prostate cancer cell lines, including PC-3, VCaP, DU-145, and LNCaP, obtaining high assay reproducibility and capture efficiencies greater than 93% after processing 7.5 mL blood samples from healthy donors, spiked with 100 cancer cells. Cancer cells remained viable after filtration due to the minimal shear stress exerted over cells during the procedure, while the identification of cancer cells by immunostaining was not affected by the number of non-specific events captured on the membrane. We were also able to identify the androgen receptor (AR) point mutation T878A from 7.5 mL blood samples spiked with 50 LNCaP cells using RT-PCR and Sanger sequencing. Finally, CTCs were detected in 8 of 8 samples from patients diagnosed with metastatic prostate cancer (mean $\pm$ SEM = 21 $\pm$ 2.957 CTCs/mL, median = 21 CTC/mL), thereby validating the potential clinical utility of the device.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。