arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1912.01255physics.app-phcond-mat.mtrl-sci
收藏

同步辐射暗场X射线显微镜实验用反射炉

Mirror Furnace for Synchrotron Dark Field X-ray Microscopy Experiments

C. Yildirim, H. Vitoux, L. E. Dresselhaus-Cooper, R. Steinmann, Y. Watier, P. K. Cook, M. Kutsal, C. Detlefs

摘要:

本文介绍了一种用于同步辐射X射线实验的多功能镜炉。该炉专门针对ESRF光束线ID 06处晶体材料的暗场X射线显微镜(DFXM)进行了优化。该炉可达到~1600°C,稳定性优于2°C,加热和冷却速率高达30°C/s。非接触式设计使样品能够在空气中或在毛细管中的受控气氛中加热。通过α-铁颗粒的热膨胀校准温度。通过将热电偶扫描通过炉的焦点来测量y轴和z轴上的温度分布。在束线的当前配置中,熔炉可用于DFXM、近场X射线形貌、明场X射线纳米层析成像、高分辨率倒易空间映射和有限的粉末衍射实验。作为第一个应用程序,我们提出了一个DFXM案例研究等温加热的商业纯铝单晶。

英文摘要:

We present a multi-purpose mirror furnace designed for synchrotron X-ray experiments. The furnace is optimized specifically for dark-field X-ray microscopy (DFXM) of crystalline materials at the beamline ID06 of the ESRF. The furnace can reach up to ~1600°C with stability better than 2°C, and heating and cooling rates up to 30°C/s. The contact-less design enables samples to be heated either in air or in a controlled atmosphere in a capillary tube. The temperature was calibrated via the thermal expansion of an a-iron grain. Temperature profiles in the y and z axes were measured by scanning a thermocouple through the focal spot of the furnace. In the current configuration of the beamline, the furnace can be used for DFXM, near-field X-ray topography, bright field X-ray nanotomography, high resolution reciprocal space mapping, and limited powder diffraction experiments. As a first application, we present a DFXM case study on isothermal heating of a commercially pure Al single crystal.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑