UPV/EHU直接发射率测量的更新测量方法和不确定度预算
Updated measurement method and uncertainty budget for direct emissivity measurements at UPV/EHU
摘要:
这项工作报告了UPV/EHU直接发射率测量设施的升级情况。仪器的改进包括高真空系统和更宽的温度范围(300-1273 K)。方法的发展,包括一个完善的测量方程更新参数和一个重新制定的ISO兼容的不确定性预算,和一个Monte Carlo程序,从光谱数据的总发射率的准确计算。这些升级已在金属和陶瓷材料的测量中得到证明和验证。在这项工作中获得的结果是适用于类似的实验装置的发射率测量,以报告可靠的发射率数据。
英文摘要:
This work reports on the upgrades made to the direct emissivity measurement facility of the UPV/EHU. The instrumental improvements consist of, among others, a high-vacuum system and a wider temperature range (300-1273 K). Methodological developments include a refined measurement equation with updated parameters and a reworked ISO-compliant uncertainty budget, and a Monte Carlo procedure for accurate calculations of total emissivities from spectral data. These upgrades have been demonstrated and validated in measurements of both metallic and ceramic materials. The results obtained in this work are applicable to similar experimental devices for emissivity measurements in order to report reliable emissivity data.