Metered reagent injection into microfluidic continuous flow sampling for conductimetric ocean dissolved inorganic carbon sensing
摘要:
持续和自主测量海洋中的总溶解无机碳(TCO 2)对于模拟重要的气候变化因素(如海洋吸收大气CO2和海洋酸化)至关重要。因此,需要小型化的化学分析系统,该系统足够小以集成到现有的Argo海洋浮子网络中,用于以实验室台式分析仪的精度对溶解的CO2进行长期无人值守的深度剖析。基于微流体电导率的方法提供了这种微生物学的可能性。>3年运行的试剂有效载荷是一个关键参数。在这里,酸的精确注入到样品中,从海水中释放CO2。制作了激光刻蚀微流控蛇形通道限流器和非对称Y流量计,用于调节海水和酸模拟物之间的计量比。改变激光蚀刻条件以产生低至约75微米的通道尺寸范围。通道流量与压力测量用于确定流体动力学阻力,并与使用一系列横截面轮廓和面积的有限元模拟进行比较。微流体计量电路由可变电阻蛇形通道和尺寸对称或不对称的Y形接头构成。样品与酸的体积比(米比)高达100:1,长度> 1米的蛇形通道宽300微米。在最高的图案分辨率下,这将需要> 600 mm 2(6 x10-4 m2)的占地面积。仅基于非对称Y形结的电路可实现高达16:1的米数比,占地面积成本< 40 mm 2,精度值约为0.2%。需要进一步的设计和制造改进,以确保在未来将计量单元集成到完整的TCO 2分析微流体回路中时这种小通道的结构和尺寸完整性。
英文摘要:
Continuous and autonomous measurement of total dissolved inorganic carbon (TCO2) in the oceans is critical for modelling important climate change factors such as ocean uptake of atmospheric CO2 and ocean acidification. Miniaturised chemical analysis systems are therefore required which are small enough for integration into the existing Argo ocean float network for long-term unattended depth profiling of dissolved CO2 with the accuracy of laboratory bench analysers. A microfluidic conductivity-based approach offers the potential for such miniaturisation. Reagent payload for >3 yr operation is a critical parameter. The precise injection of acid into sample, liberating CO2 from seawater, is addressed here. Laser etched microfluidic snake channel restrictors and asymmetric Y meters were fabricated to adjust the metering ratio between seawater and acid simulants. Laser etching conditions were varied to create a range of channel dimensions down to ~75 microns. Channel flow versus pressure measurements were used to determine hydrodynamic resistances which were compared with finite element simulations using a range of cross-section profiles and areas. Microfluidic metering circuits were constructed from variable resistance snake channels and dimensionally symmetric or asymmetric Y-junctions. Sample to acid volume ratios (meter ratio) up to 100:1 have been achieved with 300 microns wide snake channel for lengths > 1m. At the highest pattern resolution, this would require a footprint of > 600 mm2 (6 x10-4 m2). Circuits based solely on asymmetric Y-junctions gave meter ratios up to 16:1 with a footprint cost of < 40 mm2 and precision values of ~0.2%. Further design and fabrication refinements will be required to ensure the structural and dimensional integrity of such small channels in future integration of metering units into full TCO2 analysis microfluidic circuits.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。