用于血管内导航的柔性器械的光纤形状传感

Fiber Optical Shape Sensing of Flexible Instruments for Endovascular Navigation

Sonja Jäckle, Tim Eixmann, Hinnerk Schulz-Hildebrandt, Gereon Hüttmann, Torben Pätz

摘要:

目的:血管内主动脉修复术目前采用2D透视成像进行。基于光纤布拉格光栅的跟踪系统是一种新兴的技术,用于微创仪器的导航,可以减少X射线曝光和使用的造影剂。柔性结构的形状感测是具有挑战性的,并且包括许多容易产生不同误差的计算步骤。为了减少这种误差,我们提出了一种优化的形状传感模型。研究方法:我们分析了形状感测过程的每一步,可能发生哪些误差,误差如何影响形状以及如何补偿或最小化。在传感长度为38 cm的多芯光纤系统上进行了实验,分析了不同方法和参数的影响。此外,我们将3D形状重建与纤维的相应CT扫描的分割形状进行了比较,以评估我们优化的形状感测模型的准确性。最后,我们通过使用从患者数据创建的3D打印血管系统在现实的血管内场景中测试了我们的模型。结果如下:根据形状的复杂性,我们在整个38 cm的传感长度上达到了0.35至1.15 mm的平均误差和0.75至7.53 mm的最大误差。在血管内情况下,我们获得的平均和最大误差分别为1.13 mm和2.11 mm。结论:的3D形状传感模型的精度是有前途的,我们计划结合基于光纤布拉格光栅的电磁传感器系统的位置和方向的形状传感,以获得定位的导管的形状。

英文摘要:

Purpose: Endovascular aortic repair procedures are currently conducted with 2D fluoroscopy imaging. Tracking systems based on fiber Bragg gratings are an emerging technology for the navigation of minimal-invasive instruments which can reduce the x-ray exposure and the used contrast agent. Shape sensing of flexible structures is challenging and includes many calculations steps which are prone to different errors. To reduce this errors, we present an optimized shape sensing model. Methods: We analyzed for every step of the shape sensing process, which errors can occur, how the error affects the shape and how it can be compensated or minimized. Experiments were done with a multicore fiber system with 38 cm sensing length and the effects of different methods and parameters were analyzed. Furthermore we compared 3D shape reconstructions with the segmented shape of the corresponding CT scans of the fiber to evaluate the accuracy of our optimized shape sensing model. Finally we tested our model in a realistic endovascular scenario by using a 3D printed vessel system created from patient data. Results: Depending on the complexity of the shape we reached an average error of 0.35 to 1.15 mm and maximal error of 0.75 to 7.53 mm over the whole 38 cm sensing length. In the endovascular scenario we obtained an average and maximal error of 1.13 mm and 2.11 mm, respectively. Conclusions: The accuracies of the 3D shape sensing model are promising and we plan to combine the shape sensing based on fiber Bragg gratings with the position and orientation of an electromagnetic sensor system to obtain the located shape of the catheter.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑