arXiv 1908.01283physics.app-phphysics.atom-phphysics.ins-det
收藏

Sub-millimetric ultra-low-field MRI detected in situ by a dressed atomic magnetometer

Giuseppe Bevilacqua, Valerio Biancalana, Yordanka Dancheva, Antonio Vigilante

英文摘要:

Magnetic Resonance Imaging (MRI) is universally acknowledged as an excellent tool to extract detailed spatial information with minimally invasive measurements. Efforts toward ultra-low-field (ULF) MRI are made to simplify the scanners and to reduce artefacts and incompatibilities. Optical Atomic Magnetometers (OAMs) are among the sensitive magnetic detectors eligible for ULF operation, however they are not compatible with the strong field gradients used in MRI. We show that a magnetic-dressing technique restores the OAMs operability despite the gradient, and we demonstrate sub-millimetric resolution MRI with a compact experimental setup based on an in situ detection. The proof-of-concept experiment produces unidimensional imaging of remotely magnetized samples with a dual sensor, but the approach is suited to be adapted for 3-D imaging of samples magnetized in loco. An extension to multi-sensor architectures is also possible.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑