Monitoring cementless femoral stem insertion by impact

Antoine Tijou, Giuseppe Rosi, Romain Vayron, Hugues Lomami, Philippe Hernigou, Charles-Henri Flouzat-Lachaniette, Guillaume Haiat

摘要:

股骨柄(FS)植入物的初始稳定性决定了非骨水泥型髋关节置换术的手术成功率。在插入过程中,必须找到16次撞击的数量和能量的折衷方案,撞击的数量和能量应足够大,以获得FS的适当初始稳定性,而不要太高17,以降低骨折风险。本研究的目的是确定是否可以使用装有18个力传感器的锤子来监测FS的插入。19个不同尺寸的非骨水泥型FS用器械锤打入4个人工股骨,20个导致72种配置。当外科医生凭经验认为FS完全插入时,记录的撞击次数为Nsurg。使用视频运动跟踪评估插入深度E,并估计与插入结束对应的撞击22次数Nvid。对于每次撞击,基于对作为时间的函数的力的变化的分析来确定两个标记为I和D的指标23。24拔出力F与指标I显著相关(R ${}^2$ =0.67)。使用阈值来分析D的变化以确定撞击数Nd,其显示与Nsurg和Nvid密切相关,平均差异约为0.2。该方法允许确定i)外科医生应当停止压紧程序以获得FS 28的最佳插入的时刻27和ii)FS植入物的初始稳定性。这项研究铺平了道路的决策29支持系统的发展,以协助外科医生在髋关节置换术。

英文摘要:

The primary stability of the femoral stem (FS) implant determines the surgical success of cementless 15 hip arthroplasty. During the insertion, a compromise must be found for the number and energy of 16 impacts that should be sufficiently large to obtain an adapted primary stability of the FS and not too high 17 to decrease fracture risk. The aim of this study is to determine whether a hammer instrumented with a 18 force sensor can be used to monitor the insertion of FS. 19 Cementless FS of different sizes were impacted in four artificial femurs with an instrumented hammer, 20 leading to 72 configurations. The impact number when the surgeon empirically felt that the FS was fully 21 inserted was noted Nsurg. The insertion depth E was assessed using video motion tracking and the impact 22 number Nvid corresponding to the end of the insertion was estimated. For each impact, two indicators 23 noted I and D were determined based on the analysis of the variation of the force as a function of time. 24 The pull-out force F was significantly correlated with the indicator I (R${}^2$ =0.67). The variation of D was 25 analyzed using a threshold to determine an impact number Nd, which is shown to be closely related to 26 Nsurg and Nvid, with an average difference of around 0.2. This approach allows to determine i) the moment 27 when the surgeon should stop the impaction procedure in order to obtain an optimal insertion of the FS 28 and ii) the FS implant primary stability. This study paves the way towards the development of a decision 29 support system to assist the surgeon in hip arthroplasty.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑