测量法布里-珀罗干涉仪腔缺陷的新技术

New technique to measure the cavity defects of Fabry-Perot interferometers

V. Greco, A. Sordini, G. Cauzzi, K. Reardon, F. Cavallini

摘要:

(摘要):我们定义和测试一种新的技术,以准确地测量空腔缺陷的空气间隔FPI,包括由于光谱调谐过程中典型的天文观测失真。我们进一步开发了一种校正技术,以保持腔的形状尽可能恒定的光谱扫描过程中。这些是使用一个或多个FPI优化二维摄谱仪的光谱透射轮廓的必要步骤。我们设计了一个概括的技术开发的所谓的相移干涉的情况下FPI。该技术适用于任何可以通过改变腔间距($z$轴)来调谐的FPI,并且可以用于任何标准具,而不管涂层的反射率。我们的方法的主要优势是能够充分表征腔在光谱扫描,允许确定扫描相关的修改的板。作为一个测试,我们已经将这种技术应用到三个50毫米直径的干涉仪,与腔间隙范围在600微米和3毫米之间,涂层用于在可见光范围内。我们获得准确和可靠的措施,空腔缺陷的空气间隔FPI,和他们的演变过程中的整个光谱扫描。我们的主要,意想不到的,结果是,两个FPI板之间的相对倾斜显着变化,在光谱扫描,并可以主导腔缺陷,特别是,我们观察到,倾斜分量在扫描的极端是明显大于在扫描的中心。利用电子控制器的能力,以设置在任何给定的光谱步骤的参考平面,我们开发了一种校正技术,允许在一个完整的光谱扫描的倾斜最小化。这种修正在很长一段时间内保持高度稳定,远远超过了天文观测的典型持续时间。

英文摘要:

(Abridged): We define and test a new technique to accurately measure the cavity defects of air-spaced FPIs, including distortions due to the spectral tuning process typical of astronomical observations. We further develop a correction technique to maintain the shape of the cavity as constant as possible during the spectral scan. These are necessary steps to optimize the spectral transmission profile of a two-dimensional spectrograph using one or more FPIs. We devise a generalization of the techniques developed for the so-called phase-shifting interferometry to the case of FPIs. The technique is applicable to any FPI that can be tuned via changing the cavity spacing ($z$-axis), and can be used for any etalon regardless of the coating' reflectivity. The major strength of our method is the ability to fully characterize the cavity during a spectral scan, allowing for the determination of scan-dependent modifications of the plates. As a test, we have applied this technique to three 50 mm diameter interferometers, with cavity gaps ranging between 600 micron and 3 mm, coated for use in the visible range. We obtain accurate and reliable measures of the cavity defects of air-spaced FPIs, and of their evolution during the entire spectral scan. Our main, and unexpected, result is that the relative tilt between the two FPI plates varies significantly during the spectral scan, and can dominate the cavity defects; in particular, we observe that the tilt component at the extremes of the scan is sensibly larger than at the center of the scan. Exploiting the capability of the electronic controllers to set the reference plane at any given spectral step, we develop a correction technique that allows the minimization of the tilt during a complete spectral scan. The correction remains highly stable over long periods, well beyond the typical duration of astronomical observations.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑