头颈部导航干预:新的便捷领域generator的标准化评估
Navigated interventions in the head and neck area: standardized assessment of a new handy field generator
摘要:
电磁(EM)跟踪使得能够在由EM场发生器(FG)发射的磁场内定位手术器械。通常,FG越大,其跟踪体积越大。NDI(Northern Digital Inc.,Waterloo,ON,Canada)最近推出了Planar 10-11 FG,它结合了紧凑的结构(97 mm x 112 mm x 31 mm)和相对较大的圆柱形跟踪体积(直径:340 mm,高度:340 mm)。使用Hummel等人的标准化评估方案,测试了FG的跟踪精度和对外部干扰源的鲁棒性。平均位置误差(根据Hummel方案的5cm距离度量)为0.59 mm,标准设置中的平均抖动为0.26 mm。平均定向误差为0.10°。金属干扰源引起的最大位置误差(4.82 mm)是由SST 303钢引起的。相比之下,不锈钢416引起的位置误差最小(0.10 mm)。总体而言,与其他NDI FG相比,Planar 10-11 FG往往能实现更好的跟踪精度结果。由于其紧凑的结构和便携性,FG可能有助于增加EM跟踪系统的临床使用。
英文摘要:
Electromagnetic (EM) tracking enables localization of surgical instruments within the magnetic field emitted by an EM field generator (FG). Usually, the larger a FG is, the larger its tracking volume is. However, the company NDI (Northern Digital Inc., Waterloo, ON, Canada) recently introduced the Planar 10-11 FG, which combines a compact construction (97mm x 112mm x 31mm) with a relatively large, cylindrical tracking volume (diameter: 340mm, height: 340mm). Using the standardized assessment protocol of Hummel et al., the FG was tested with regard to its tracking accuracy and to its robustness with respect to external sources of disturbance. The mean positional error (5cm distance metric according to Hummel protocol) was 0.59mm, with a mean jitter of 0.26mm in the standard setup. The mean orientational error was found to be 0.10°. The highest positional error (4.82mm) due to metallic sources of disturbance was caused by the steel SST 303. In contrast, steel SST 416 caused the lowest positional error (0.10mm). Overall, the Planar 10-11 FG tends to achieve better tracking accuracy results compared to other NDI FGs. Due to its compact construction and portability, the FG could contribute to increased clinical use of EM tracking systems.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。