偏振分束器的无对准特性

Alignment-free characterization of polarizing beamsplitters

Felipe Calliari, Pedro Tovar, Christiano Nascimento, Breno Perlingeiro, Gustavo Amaral, Guilherme Temporão

摘要:

用于测量偏振分束器(PBS)参数(尤其是消光比)的传统方法需要高度偏振光源、对准过程和/或随时间变化的实验参数(例如偏振旋转)。在这项工作中,提出了一种新的方法,它采用非偏振光和法拉第镜。结果表明,精确的消光比和插入损耗值可以实现在三个单扫描测量,没有任何对准要求或任何种类的随时间变化的信号。

英文摘要:

Traditional methods for measurement of Polarizing Beamsplitter (PBS) parameters, especially the extinction ratio, require highly polarized light sources, alignment procedures and/or experimental parameters that change over time, such as polarization rotations. In this work, a new method is presented, which employs unpolarized light and a Faraday Mirror. It is shown that precise extinction ratio and insertion loss values can be achieved in three single-sweep measurements, without any alignment requirements or time-varying signals of any kind.