径向速度变化测量用ThNe校准灯的最佳工作条件

On the optimum operating conditions of ThNe calibration lamps for measurements of radial velocity variations

M. Ammer-von Eiff, E. W. Guenther

摘要:

以前的研究表明,ThNe灯非常适合于波长变化的精确测量。所有空心阴极灯(HCL)的一个关键问题是灯的工作电流-低值的优点是灯的寿命更长,而有用的线的数量更少。我们调查的合适的谱线的数量取决于当前使用的和获得的光谱的ThNe HCL耦合由一根纤维的陶登堡Echelle摄谱仪在一个设置,这是典型的许多高分辨率摄谱仪。在制造商规定的范围内选择工作电流。我们改变了曝光时间,以确定饱和度和噪声的影响。在所考虑的波长范围内自动识别了数千条Th线。当改变设置时,我们注意到Th线的几百ms $^{-1}$的分散,但没有检测到任何系统性趋势。然而,Th线的波长残差的分散表明,可能需要精确控制工作电流。个别线的校准精度的估计表明,强Th线的组合允许达到1 $\mathrm{ms}^{-1}$的水平。尽管高工作电流会缩短灯的寿命并加速其老化,但由于与Ne线相比,大量Th线变得更强,因此它可以保证最高的精度。然后,通过调整曝光时间,可以最小化噪声、饱和度、所识别的线的波长残差以及由于饱和Ne线造成的污染。我们的研究结果表明,ThNe灯比ThAr灯更适合于各种研究,涉及径向速度变化的精确测量,例如晚型恒星,棕矮星,遭受强烈消光的物体,或高红移星系的研究。

英文摘要:

Previous studies have shown that ThNe lamps are very suitable for the precise measurement of wavelength variations. A critical issue for all hollow cathode lamps (HCL) is the current at which the lamp is operated - a low value has the advantage that the life time of the lamp is longer whereas the number of useful lines is lower. We investigate the number of suitable spectral lines depending on the current used and obtained spectra of a ThNe HCL coupled by a fibre to the Tautenburg Echelle spectrograph in a setting that is typical for many high-resolution spectrographs. Operating currents were chosen in the range specified by the manufacturer. We varied exposure time to identify the saturation level and effects of noise. A few thousands of Th lines were identified automatically in the wavelength range considered. We noticed a scatter of several hundred ms$^{-1}$ of Th lines when varying the settings but did not detect any systematic trends. The scatter in wavelength residuals of Th lines however indicates that a precise control of operating current can be necessary. An estimation of the calibration precision of individual lines indicates that a combination of strong Th lines allows one to reach a level of 1$\mathrm{ms}^{-1}$. Although a high operating current reduces the life-time of the lamp and accelerates its ageing it guarantees highest precision because of the numerous Th lines which become strong compared to the Ne lines. Then, noise, saturation, wavelength residuals of lines identified, and the pollution due to saturated Ne lines can be minimised by adjusting exposure time. Our results indicate that ThNe lamps can be better suited than ThAr lamps for all kinds of studies that involve the precise measurement of radial velocity variations, e.g. studies on late-type stars, brown dwarfs, objects that suffer from strong extinction, or high-redshift galaxies.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑